LDS 3000

 

Air blower

Purging of CD 3002 using air blower

1

1.

Purging of CD 3002 using air blower

1.1

General

In application with high dust load and ash build-up in front of the lens in 
CD 3002 very high purging flows may be necessary. For these applica-
tions an air blower producing flow rates in the 1000 ltr/min. range can be 
used. A typical application where this is preferable is in MWI (municipal 
waste incineration). This manual relates to the installation and operation 
of the air blower set supplied by Siemens Laser Analytics.

1.2

Installation

An important issue is the counter pressure build up at the output of the 
air blower. It is vital to keep this as low as possible since the air flow is 
decreasing when the counter pressure increases – se the graph below.

Flow rate vs counter pressure.

A large cross section of the air blower output and a short tube is param-
eters that affects this in a positive direction. If a long tube must be used 
– make sure it has a large diameter. The following diagram illustrates the 
flow changes with different output tubes and may serve as a guide when 
the set-up is designed. If the air blower is serving two sensors it is very 
important that the lengths of the tubes to the two sensors are equal. This 
ensures that the air is split 50/50 between the two.

0

200

400

600

800

1000

0

40

80

120

160

200

Flow rate [ltr/min]

Counter pressure [mbar]

Flow rate vs counter pressure

Summary of Contents for CD 3002

Page 1: ...Your eye in hell Air Blower for CD 3002 ...

Page 2: ... Siemens Laser Analytics AB Jan 2001 CD 3002 Air blower manual Document number CID 3002 1110 Rev 1A ...

Page 3: ...LDS 3000 Air blower Content I Content 1 Purging of CD 3002 using air blower 1 1 1 General 1 1 2 Installation 1 1 3 Maintanance 4 ...

Page 4: ...Content LDS 3000 Air blower II ...

Page 5: ...at the output of the air blower It is vital to keep this as low as possible since the air flow is decreasing when the counter pressure increases se the graph below Flow rate vs counter pressure A large cross section of the air blower output and a short tube is param eters that affects this in a positive direction If a long tube must be used make sure it has a large diameter The following diagram i...

Page 6: ...e air blower and the sensor Mechanical The air blower should be monted close to the sensor using bolts of the di mension M10 on the following positions Position of mounting holes 1 10 100 1000 Length of tube m Air flow ltr min Maximum flow Q m Maximum flow Q 900 ltr min 9 m n 900 ltr min 226 mm 220 mm ...

Page 7: ...lower should either be horizontal with the mount ing plate facing downwards or vertical with the air exit facing upwards The figure below shows a good way of setting up the sensor with an air blower Example set up of CD 3002 with air blower Output air tube Input air filter Air blower ...

Page 8: ...e the input filter at a regular interval The time between changes is totally dependent on the environment in which the air blower operates The inlet under pressure directly after the filter will reveal the condition of the filter and can be used to monitor this Siemens Laser Analytics can provide suitable equippment for this ...

Reviews: