Siemens AG
SPR2-330.815.01.03.02
ARCADIS Avantic
12.06
CS PS SP
General Information
15
Page 15 of 66
Medical Solutions
Fig. 2:
Measuring circuit for measuring the system leakage current according to the differential
current method in compliance with DIN VDE 0751-1/2001-10, Fig. C6 for protection
class I.
Pos. 1
= System
Pos. 2
= Application part type B (if available)
Pos. 3
= Measurement setup (integrated into measuring device)
If the direct measurement of the leakage current is used (measurement setup according
to
), the system must be insulated during the measurement and must not be
touched.
Fig. 3:
Measuring circuit for direct measurement of the system leakage current in compliance