System Misinterprets Wash 1 Fluid Signals 

If the bottle is replaced while in the Ready state, you should prime the line using the 
following steps: 

a. 

Select the system icon in the top left of the workspace window. 

b. 

From the menu list, select 

Diagnostic Tools

The System – Diagnostic Tools window displays. 

c. 

Under Prime Functions select 

Prime System Fluid Bottle

 from the list. 

d. 

From the Fluid list, select 

Wash 1

e. 

Select 

Perform

 

If the Wash 1 bottle becomes empty and the Supplies Status button does not display yellow 
to warn you that Wash 1 is low, call your local support representative to have a Field Service 
Engineer visit your site and inspect the system. 

System Stops Processing Samples When Wash 1 is Present 

Normally, if the Wash 1 bottle is depleted but the Wash 1 reservoir contains fluid, the Supplies 
Status button on the UI workspace displays yellow indicating that Wash 1 is “Present, using 
reserve”. If, after you replace the empty Wash 1 bottle, the Supplies Status button continues to 
display yellow, or if there was already sufficient Wash 1 in the bottle, perform the following steps 
to attempt to clear the status or error message: 

The following steps may not clear the status or error message if the Wash interconnect PCB is 
malfunctioning or if there is another system problem. Please call your local support 
representative to report the problem after attempting to clear the status or error message. 

 

Clearing a Yellow Supplies Status Button 

a. 

Lift the lid of the Wash 1 bottle.  

b. 

Close the lid of the Wash 1 bottle. 

The system displays “Did you refill the Wash 1 Bottle?”  

c. 

Select 

Yes

.  

d. 

Prime the Wash 1 fluid bottle following instructions in the Operator’s Guide. 

e. 

Call your local support representative and notify them that the Supply Status button 
continues to display yellow when sufficient Wash 1 is in the bottle. 

 

Clearing a Red Error Message: 

Normally, after the Wash 1 bottle and reservoir are depleted the system issues a red 
“Reservoir depleted” and “Sampling Stopped” message. If the red error message continues 
to display after you replace an empty Wash 1 bottle, or if there was already sufficient  
Wash 1 in the bottle, then perform the following:  

a. 

Prime the Wash 1 fluid lines following instructions in the Operator’s Guide. 

b. 

Refer to the ADVIA Centaur XP Operator’s Guide or ADVIA Centaur Operator’s Guide 
for additional instructions. 

c. 

Call your local support representative and notify them that the red message displays 
when sufficient Wash 1 is present. 

Urgent Field Safety Notice 10813924, Rev. A – November 2012 

Page 3 of 5

GPF-003-11 V1.0 Effective: September 21, 2012 
Related Procedure:  GP-003 DX- Field Correction Action 

 

Summary of Contents for ADVIA Centaur

Page 1: ...eservoir is empty when it is actually full If the system fails to detect an empty Wash 1 bottle and reservoir it can continue to process samples and produce erroneous but believable results which are falsely high or low depending if the assay architecture is direct or indirect This issue potentially affects only those assays that use Wash 1 Table 1 Table 1 Assays Potentially Affected Assay Type As...

Page 2: ...age indicating that fluid is depleted and the system is going to stop sampling This case has no affect on results Actions to be Taken by the Customer Please use the actions below to address the following issues 1 The system fails to detect that the Wash 1 bottle and reservoir are empty and continues to process samples 2 The system stops processing samples even though there is sufficient Wash 1 pre...

Page 3: ...nterconnect PCB is malfunctioning or if there is another system problem Please call your local support representative to report the problem after attempting to clear the status or error message Clearing a Yellow Supplies Status Button a Lift the lid of the Wash 1 bottle b Close the lid of the Wash 1 bottle The system displays Did you refill the Wash 1 Bottle c Select Yes d Prime the Wash 1 fluid b...

Page 4: ...e who may have received this product We apologize for the inconvenience this situation has caused If you have any questions please contact your Siemens Technical Solutions Center or your local Siemens technical support representative ADVIA Centaur is a trademark of Siemens Healthcare Diagnostics Urgent Field Safety Notice 10813924 Rev A November 2012 Page 4 of 5 GPF 003 11 V1 0 Effective September...

Page 5: ... of the enclosed Siemens Healthcare Diagnostics Urgent Field Safety Notice dated November 2012 regarding System Misinterprets Wash 1 Fluid Signals 10813924 Please read each question and indicate the appropriate answer Fax this completed form to Siemens Healthcare Diagnostics at the fax number indicated at the bottom of this page I have read and understood the Urgent Field Safety Notice instruction...

Reviews: