Siemens ACUSON X150 User And Reference Manual Download Page 11

S u p p l e m e n t   t o   U s e r   a n d   R e f e r e n c e   M a n u a l s  

1 1   /   18 

To: Do 

This: 

Create a network profile 

1. Click 

New

The ultrasound system displays the 

Network Profile Settings

 dialog box. 

2.  Enter the required network profile settings and then click 

OK

The ultrasound system displays the new network profile. 

Change network profile settings 

1.  Select a wireless network. 
2. Click 

Edit

3.  Enter the required network profile settings and then click 

OK

4.  To delete the existing profile setting, click 

Remove

Enable password protection to 
restrict changes to the wireless 
network options 

1. Click 

Lock

2.  Enter a new password and then click 

OK

Disable password protection to 
provide unrestricted access to the 
wireless network options 

1. Click 

Unlock

2.  Enter the password and then click 

OK

Change a password 

1. Click 

Change

2.  Enter the password. 
3. Reenter the new password and then click 

OK

Display characters of a password 
as you enter the password 

 Select 

Show password

Ultrasound System Security (Virus Protection) 

Ultrasound System Security is virus protection software that Siemens recommends for 
ultrasound systems connected to a LAN or wireless network. To enable Ultrasound System 
Security, contact your Siemens service representative. 

Summary of Contents for ACUSON X150

Page 1: ...Siemens Medical Solutions USA Inc 10658982 02 10658982 ABS 001 02 02 BZSB 1 18 ACUSON X300 ACUSON X300 PE ACUSON X150 Supplement to the User and Reference Manuals ...

Page 2: ... marking only applies to medical devices that have been put on the market according to the above referenced Council Directive Unauthorized changes to this product are not covered by the CE marking and the related Declaration of Conformity EU Authorized Representative Siemens AG Medical Solutions Henkestrasse 127 91052 Erlangen Germany Global Business Unit and Legal Manufacturer Siemens Medical Sol...

Page 3: ...Network 10 Ultrasound System Security Virus Protection 11 Viewing Patient Reports with Qview 12 Report Option Configuration Selections 13 QuickStart Battery Powered Standby Mode 14 BP9 4 Transducer Imaging 14 Imaging with the BP9 4 Transducer 14 Updates to the System Controls 15 Needle Guide Brackets 16 BP9 4 Disposable Endocavity Biopsy Straight Needle Guide Kit 1 16 BP9 4 Disposable Endocavity B...

Page 4: ...manuals for the devices listed within this publication This supplement provides new information about features accessories settings and standards Refer to the user and reference manuals for a complete set of operating instructions for your ultrasound system Compatible Systems and Product Versions ACUSON X300 PE 7 5 ACUSON X300 2 5 SONOVISTA X300 2 5 7 5 ACUSON X150 2 5 ...

Page 5: ...io interference from the 2 4 GHz frequency bandwidth Approved Wireless Adapters Caution Use of wireless adapters not tested for use and approved by Siemens can damage the ultrasound system Do not connect any wireless adapters to the system except the wireless adapters listed below Note Use of an unapproved wireless adapter may cause loss of data transferred over a wireless network The following wi...

Page 6: ...sceptible to electromagnetic interference generated by other RF energy sources To prevent this interference increase the distance between the ultrasound system and the interfering RF energy source Siemens recommends connecting the wireless adapter to the leftmost USB port on the Input Output panel of the ultrasound system Informational Icons for the Wireless Adapter Icon Explanation Connected to a...

Page 7: ...ox to configure the network profile The ultrasound system uses the network profile to connect to the selected wireless network See also Network Profile Settings p 8 Connect Connect to the selected wireless network Disconnect Disconnect from the selected wireless network Scan Refresh the list of available wireless networks Connection Information SSID Service Set Identification View the name of the ...

Page 8: ...Type None WEP TKIP AES Select the encryption method for the connection to a wireless network Note The options for the encryption method are dependent on the wireless network Security Key text entry Enter the security key for the connection to a wireless network EAP Authentication Note EAP authentication options are enabled only when you select WPA or WPA2 for the security type EAP Type PEAP Select...

Page 9: ...se the following IP address IP Address text entry Specify a static IP address Subnet Mask text entry Specify the subnet mask for a LAN Local Area Network Default Gateway text entry Specify the gateway for the LAN Obtain DNS server address automatically on off Configure the ultrasound system to use the address of the DNS Domain Name Server sent by the DNS service on or to use the address specified ...

Page 10: ...red networks to ensure data security Select Enabled Deactivate the connection to a wireless network Select Disabled Connect to the selected network 1 Click Scan to refresh the list of wireless networks Note The wireless adapter must remain connected to the ultrasound system while refreshing the list 2 Select a network from the list 3 Click Connect 4 If the system displays the Network Profile Setti...

Page 11: ...ion to restrict changes to the wireless network options 1 Click Lock 2 Enter a new password and then click OK Disable password protection to provide unrestricted access to the wireless network options 1 Click Unlock 2 Enter the password and then click OK Change a password 1 Click Change 2 Enter the password 3 Reenter the new password and then click OK Display characters of a password as you enter ...

Page 12: ...e patient report using the layout provided by Qview 2 Press Report to exit the report page Or for the ACUSON X300 PE system and the ACUSON X150 system press the Exit soft key View the patient report using the layout provided by Qview and the patient report layout provided by the ultrasound system Prerequisite The Both selection must be enabled in the system presets 1 Press Report The system displa...

Page 13: ...ystem Report Only enables only the layout provided by the ultrasound system Telexy Qview Only enables only the layout provided by Qview Both enables the layouts provided by the ultrasound system and by Qview Note The patient report provided by the ultrasound system displays by default when you press Report Login Information Address text Enter the address of the Qview server User Name text Enter th...

Page 14: ... 2D imaging with the BP9 4 transducer Live dual format displays both the transverse image and the sagittal image in real time When you have acquired an image you can change the orientation of the image make measurements and calculations annotate with pictograms or text or perform a biopsy or puncture procedure Note During live dual format you cannot change the orientation of the image To Do This S...

Page 15: ...ual format Press the STEER control on the control panel S T Example of live dual format before the sagittal image is repositioned S T Example of live dual format when the STEER control is pushed to the left T S Example of live dual format when the STEER control is pushed to the right Updates to the System Controls Supplement to the Instructions for Use Appendix B page B 15 For the BP9 4 transducer...

Page 16: ...s for attachment and care procedures including cleaning sterilization and disposal BP9 4 Disposable Endocavity Biopsy Straight Needle Guide Kit 2 The disposable endocavity needle guide is a single use item Refer to the in box instructions for attachment and care procedures including cleaning sterilization and disposal BP9 4 Reusable Endocavity Biopsy Straight Needle Guide The stainless steel endoc...

Page 17: ...only the following three transducers for THI Imaging Functions Technical Description Appendix A Instructions for Use the frequencies for other compatible transducers are unchanged CH5 2 1 5 MHz 1 8 MHz 2 2 MHz C8 5 3 1 MHz 3 6 MHz P5 1 1 5 MHz 1 7 MHz for TCD exam only P5 1 1 7 MHz 2 0 MHz 2 2 MHz for Cardiac and Ped Echo exams P5 1 1 7 MHz 2 0 MHz for Abdomen exam only ...

Page 18: ... and IEC 60601 1 2nd and 3rd Ed EN 60601 1 1 and IEC 60601 1 1 EN 60601 1 2 and IEC 60601 1 2 EN 60601 1 4 and IEC 60601 1 4 EN 62304 and IEC 62304 EN 60601 1 6 and IEC 60601 1 6 EN 62366 and IEC 62366 EN 60601 2 18 and IEC 60601 2 18 EN 60601 2 25 and IEC 60601 2 25 EN 60601 2 37 and IEC 60601 2 37 EN ISO 14971 EN ISO 10993 1 Acoustic Output Standards IEC 61157 Declaration of Acoustic Power AIUM ...

Reviews: