9 eSieFusi on Im aging
Instructions for Use
9 - 27
Attaching the Sensors to the Electronics Unit
Ports for sensors on the electronics unit are associated with a specific transducer port on the
ultrasound system.
Sensor Port
Transducer Port
1
Left
2
Center
3
Right
Port 4 on the electronics unit is reserved for the needle sensor.
To connect a sensor to the electronics unit:
Insert the sensor connector in the corresponding port on the electronics unit until you hear
an audible click.
Attaching Sensors to the Transducer or Needle
Refer to the manufacturer's instructions included with the device.
Troubleshooting the Tracking System
An LED on the front panel of the electronics unit provides status information for troubleshooting
the tracking system. The LED slowly blinks green when the tracking system is fully functional
and the transmitter is not activated. The LED displays solid green when the tracking system is
fully functional and the transmitter is activated.
Symptom
Possible Cause
Recommend Action
Front panel LED is not
illuminated
No power.
Check the power cord connection to
the electronics unit and to the power
source (outlet).
Cycle power to the electronics unit.
Front panel LED blinks red
and yellow
A valid transmitter is not connected to
the electronics unit or is not connected
properly.
Attach a transmitter to the electronics
unit.
Check the connection from the
transmitter to the electronics unit.
Front panel LED slowly blinks
green and orange
The tracking system has not fully
initialized.
Wait a few minutes for the tracking
system to complete its initialization
process. The LED will then slowly
blink green.
The tracking system is unable
to communicate using the
USB connection
USB cable is not properly connected.
Disconnect and reconnect the USB
cable to the electronics unit.
Cycle power to the electronics unit.
Summary of Contents for Acuson S2000
Page 12: ...1 Introduction 1 2 Instructions for Use ...
Page 14: ...1 Introduction 1 4 Instructions for Use System Review Example of the ultrasound system ...
Page 84: ...2 Safety and Care 2 54 Instructions for Use ...
Page 86: ...3 System Setup 3 2 Instructions for Use ...
Page 112: ...3 System Setup 3 28 Instructions for Use ...
Page 114: ...4 Examination Fundamentals 4 2 Instructions for Use ...
Page 144: ...5 Transducer Accessories and Biopsy 5 2 Instructions for Use ...
Page 196: ...7 Specialty Transducers 7 2 Instructions for Use ...
Page 200: ...7 Specialty Transducers 7 6 Instructions for Use ...
Page 202: ...8 Physiologic Function 8 2 Instructions for Use ...
Page 208: ...9 eSieFusion Imaging 9 2 Instructions for Use ...
Page 236: ...10 Virtual Touch Applications 10 2 Instructions for Use ...
Page 258: ...10 Virtual Touch Applications 10 24 Instructions for Use ...
Page 302: ...Appendix A Technical Description A 44 Instructions for Use ...
Page 326: ...Appendix B Control Panel and Touch Screen B 24 Instructions for Use ...
Page 328: ...Appendix C Control Panel C 2 Instructions for Use ...
Page 394: ...Appendix D On screen Controls D 50 Instructions for Use ...
Page 444: ...Appendix F Acoustic Output Reference F 2 Instructions for Use ...
Page 516: ...Appendix F Acoustic Output Reference F 74 Instructions for Use ...
Page 517: ......
Page 518: ......