Powering up the System

1. To power on the system, open the LCD lid. 

2. There is a short delay from opening the lid until 

the system is ready to image.

3. Power off the system by closing the LCD lid or 

pressing 

MENU

and selecting 

Shutdown

.

Note:

The system clears the current study or patient data when powering

off. To save the information press 

STORE

.

Beginning an Exam

To register a patient:

1. Press the 

NEW PATIENT

button to enter patient 

name and ID. Use the QWERTY keypad to 

enter information.

2. The 

ENTER

or 

SELECT

keys move between fields.

3. Press the 

RETURN

or 

SELECT

keys to save the new 

patient information.

a. Press

ESC

or 

NEW PATIENT

keys to exit the new 

patient screen without saving the data.

4. Select 

EXAM TYPE.

Bypass of patient registration:

1. Begin exam.

2. Press 

STORE

to store images and loops.

3. System stores images and loops to “Unknown” 

patient identifier.

Image Screen Display:

Active exam type, depth, gain, frames per second, 

grayscale map and compression curve display in upper

left of the image screen.

If TGO™ tissue grayscale optimization technology 

is enabled, TGO gain is displayed.

Emergency Workflow

1. Open lid to power up.

2. Do not register patient, unfreeze to begin scanning.

3. Press Store to save images and loops.

4. Patient registered as “Unknown”.

Note: 

To change system screen brightness (different ambient lighting

situations) press 

MENU

and scroll down to 

SCREEN BRIGHTNESS

.

Acquiring an Image

1. The P10 system powers up in 

FREEZE

mode to 

conserve battery life.

2D Imaging:

1. Press 

FREEZE

to enter 2D-mode.

2. Use the Navigation ring buttons to make 

adjustments to depth and gain.

a. Up and down to adjust depth.

b. Left and right to adjust gain.

3. Press 

SELECT

to activate the imaging pop-up 

menu and access other imaging controls (buttons 

move up and down menu.

4. Press 

SELECT

again to activate the menu item 

and make selection.

To...

…do this:

Icon

Change image

Select 

Gain +/-

or  

gain/brightness

press up/down buttons

on navigation ring

Adjust depth

Select 

Depth +/-

or 

press up/down buttons 

on navigation ring

Activate/deactivate TGO

Select 

TGO On/Off

(2D Menu)

Display annotations

Select 

Annotations

(2D Menu)

2D Imaging Pop-up Menu:

TGO – 

Tissue Gain Optimization, or automatic 

gain/brightness uniformity and contrast resolution

adjustment. 

Zoom – 

magnifies image.

RGB Map –

selects different grey scale (RGB) 

display maps.

Depth – 

changes depth setting.

Gain – 

changes overall gain setting.

Comp Map – 

displays compression curves (changes

dynamic range of image).

Note:

TGO technology affects the whole image including artifacts and

defects. For optimal TGO performance, ensure full transducer contact,

minimize depth, and avoid bright reflectors (gas). 

Performing Measurements

Press the 

MEAS

key to the right of the Nav ring to

display the measurement window.  

To...

…do this:

Icon

Select the desired  

Select 

Caliper1

,

Caliper2

measurement

or 

Ellipse

from menu

Use calipers

Move highlighted endpoint 

with the navigation ring,  

use 

SELECT

to highlight 

second caliper

Ellipse

Press 

SELECT

to toggle 

between ellipse functions; 

size, pan or rotate 

Change Ellipse functions

Use navigation ring buttons 

to change size, rotate ellipse, 

or pan

Clear/Delete

Press 

Measurement

and select 

Clear

. Removes currently 

displayed measurements

Lock Measurements

Press 

MEASUREMENT

. Locks value 

and location and re-displays 

measurement menu. 

Store

Press 

STORE

Annotations

1. In 2D, freeze, or when recalling an image, 

press 

SELECT

.

2. Select 

Annotations

from menu, and then 

select an annotation from the list. 

System displays a highlighted box.

3. Press 

SELECT

to position the annotation 

on the screen where the highlighted 

box displays.

4. To delete annotations, select 

Clear All

from 

annotations menu.

Storing an Image

The P10 system stores still frames or loops to a

removable SD Memory Card.

1. Press 

STORE

key of frozen image to store a 

single frame.

2. Press 

STORE

while live imaging or playing a 

loop to store a loop.