ACUSON Cypress™ Operator’s Manual
5-23
2D Scanning
2D
Soft Windows
Table 16 describes how Soft Windows and their corresponding knobs on the
keyboard operate in
2D
scanning. The functions available to you depend on
which options are installed on your Cypress system.
TABLE 16.
2D
Scanning Soft Windows Options
Window/Function
Description
2D
Steer
(Linear)
Changes the scanning
angle of the
2D
image
There are three positions: 0°, +20°, –20°. Press the
corresponding Soft Window knob to select
2D STEER
.
To steer the beam to the right, turn the knob
clockwise.
To steer the beam to the left, turn the knob coun-
terclockwise.
Protocol
(when in a stress pro-
tocol)
Displays the
Stage
and
View
labels. Continue turn-
ing the knob until either
Rest
(for 2-Stage and 3-
Stage) or
Base
(for Pharmacological) is highlighted.
Depth
Adjusts the scanning
field of view
Scanning depth is measured in centimeters and
varies according to the transducer frequency. Press
the corresponding Soft Window knob to select
DEPTH
.
For a
shallower
depth, turn the knob counter-
clockwise.
For a
deeper
depth, turn the knob clockwise.
The current selection is displayed on the top left
corner of the screen (see “Cypress System Data”
on page 5-9).
Zoom
(Available only in active
scanning)
Allows you to select
and enlarge a portion
of the image.
Press the corresponding Soft Window knob to
select
ZOOM
. This places the "zoom" box on the
image. Use the Trackball to position the Zoom box
To change the magnification factor, turn the knob
clockwise.
Press <
ENTER
> to enlarge the image.
To reposition or resize the zoom box, turn the
knob counterclockwise.
Press the
ZOOM
Soft Window knob to exit.