3 S y s t e m C o n t r o l s
3 - 26
[ 1 ] I N S T R U C T I O N S F O R U S E
Image Task Card
The
Image
task card is used for selecting the imaging mode and the
image format and for optimizing the settings such as dynamic range,
persistence, and edge enhancement or for selecting an exam type or
transducer that contains pre-configured, optimized settings.
The
Image
task card also provides controls for functions other than imaging
parameters. VCR recording, ECG settings, image formatting, and the update
styles settings are organized in group boxes.
The
Image
task card also includes capture definitions for storing and
reviewing clips.
Parameter Menu Selections
Mode-specific menus and selections display on the left side of the screen in
the
Image
task card. Each imaging mode has a mode-dependent Parameter
menu used to adjust imaging parameters.
Note:
A Parameter menu may have two pages of selections. To access the menu
selections, roll the trackball to the
Page 1 of 2
indicator or the
Page 2 of 2
indicator on the
Parameter menu and then press the
SELECT
key.
Note:
A menu selection can be an icon and/or text. The "Tool Tip" column in the following
tables indicates additional text displayed on the screen when you roll the trackball to
position the trackball pointer on the menu selection. The "Settings" column in the
following tables lists the settings that are displayed on the Parameter menu for each
menu selection (graphically depicted settings or options are not listed in this column).
2D-Mode Parameter Menu Selections
Menu
Selection
Tool Tip
Description
Settings
(current
setting)
Transmit
Frequency
Changes the transmit frequency of an active
multi-frequency transducer in 2D-mode.
Transducer-
dependent
#Foc
# Focal Points
Selects the number of focal zones.
Dependent on
transducer and
exam type
DR Dynamic
Range
Controls the overall contrast resolution of the image.
30 dB to 70 dB in
5 dB increments
R/S
Res/Speed
Adjusts the balance between the image line density
(resolution) and the frame rate.
0 to 5
Edge
Edge Enhance
Distinguishes the contours of a structure during
real-time imaging.
0 (none), 1 to 3
[1] Instructions for Use
Protocol Group Box 3-38
Summary of Contents for ACUSON Antares
Page 2: ......
Page 4: ...ii 1 I N S T R U C T I O N S F O R U S E ...
Page 6: ...iv 1 I N S T R U C T I O N S F O R U S E ...
Page 10: ...viii 1 I N S T R U C T I O N S F O R U S E ...
Page 12: ...1 Introduction 1 2 1 I N S T R U C T I O N S F O R U S E ...
Page 180: ...4 System Setup 4 36 1 I N S T R U C T I O N S F O R U S E ...
Page 182: ...5 Beginning an Examination 5 2 1 I N S T R U C T I O N S F O R U S E ...
Page 200: ...5 Beginning an Examination 5 20 1 I N S T R U C T I O N S F O R U S E ...
Page 230: ...6 Technical Description 6 30 1 I N S T R U C T I O N S F O R U S E ...