3 S y s t e m C o n t r o l s
[ 1 ] I N S T R U C T I O N S F O R U S E
3 - 51
Compose Task Card
The
Compose
task card allows you to edit data acquired with system
feature options. It provides tools for manipulating or rotating the
images/volumes. When using SieScape or 3-Scape, the ultrasound system
will default to the
Compose
task card when acquisition is completed. To
perform measurements or to review images, select the
Calcs
or
Review
task cards.
During 3-Scape imaging and
four
Sight 4D imaging, the system also displays
the
Protocol
group box on the
Compose
task card.
SieScape Parameter Menu Selections
(Option)
Menu Selection
Description
Resize Fit
(Resize Image)
Full
scales the image to its full acquisition size.
1
,
2
,
3
,
4
, and
5
incrementally scale the image between
Best Fit
and
Full
.
Redisplay
Redisplay the SieScape image in the size and rotation selected prior to entering CINE.
Restore
Scales the SieScape image to the original display of the frozen image.
Best Fit
Automatically scales the image to fit the image area.
DR
(Dynamic Range)
Controls the overall contrast resolution of the image.
Maps
Selects a processing curve that assigns echo amplitudes to gray levels.
Color SieScape Parameter Menu Selections
(Option)
The selections are available during the Setup and Acquire processes. Use
these selections to optimize and acquire a Color SieScape image.
Menu Selection
Description
AllPwr
Activates or deactivates the Power display in the ROI. When
On
, Power data is
visible in the real-time image. When
Off
, Power data is captured and displays in the
SieScape image.
Color Cap
Enables continuous capture of Power data without transducer movement.
Adds the Power component even when there is insufficient motion for acquiring a
black and white SieScape image.
Flow
Activates or deactivates Power mode without stopping image acquisition.
AcqFrac
Adds either 100 percent of the Power ROI data (
On
) or 30 percent of the Power ROI
data (
Off
) to the SieScape image during SieScape image acquisition.
[1] Instructions for Use
Protocol Group Box 3-38
Summary of Contents for ACUSON Antares
Page 2: ......
Page 4: ...ii 1 I N S T R U C T I O N S F O R U S E ...
Page 6: ...iv 1 I N S T R U C T I O N S F O R U S E ...
Page 10: ...viii 1 I N S T R U C T I O N S F O R U S E ...
Page 12: ...1 Introduction 1 2 1 I N S T R U C T I O N S F O R U S E ...
Page 180: ...4 System Setup 4 36 1 I N S T R U C T I O N S F O R U S E ...
Page 182: ...5 Beginning an Examination 5 2 1 I N S T R U C T I O N S F O R U S E ...
Page 200: ...5 Beginning an Examination 5 20 1 I N S T R U C T I O N S F O R U S E ...
Page 230: ...6 Technical Description 6 30 1 I N S T R U C T I O N S F O R U S E ...