SITRANS P300 (7MF8.23.. 7MF8.24.. 7MF8.25..)
A5E03434764-01, 08/2011
25
Disposal
Devices identified by this symbol may not be disposed of in the municipal
waste disposal services under observance of the Directive 2002/96/EC on
waste electronic and electrical equipment (WEEE).
They can be returned to the supplier within the EC or to a locally approved
disposal service. Observe the specific regulations valid in your country.
NOTICE
Special disposal required
The device includes components that require special disposal.
● Dispose of the device properly and environmentally through a local waste disposal contractor.
Technical data
Input
Gauge pressure input
HART
PROFIBUS PA or Foundation Fieldbus
Measured variable
Gauge pressure
Measuring
span
Maximum
operating
pressure
MAWP (PS)
Maximum test
pressure
Measuring
range
Maximum
permissible
operating
pressure
Maximum test
pressure
0.01 …
1 bar g (0.15
… 14.5 psi g)
4 bar g
(58 psi g)
6 bar g
(87 psi g)
1 bar g
(14.5 psi g)
4 bar g
(58 psi g)
6 bar g
(87 psi g)
0.04 …
4 bar g (0.58
… 58 psi g)
7 bar g
(102 psi g)
10 bar g
(145 psi g)
4 bar g
(58 psi g)
7 bar g
(102 psi g)
10 bar g
(145 psi g)
0.16 …
16 bar g (2.3
… 232 psi g)
21 bar g
(305 psi g)
32 bar g
(464 psi g)
16 bar g
(232 psi g)
21 bar g
(305 psi g)
32 bar g
(464 psi g)
0.63 …
63 bar g (9.1
… 914 psi g)
67 bar g
(972 psi g)
100 bar g
(1450 psi g)
63 bar g
(914 psi g)
67 bar g
(972 psi g)
100 bar g
(1450 psi g)
1.6 …
160 bar g
(23.2 …
2321 psi g)
167 bar g
(2422 psi g)
250 bar g
(3526 psi g)
160 bar g
(2321 psi g)
167 bar g
(2422 psi g)
250 bar g
(3626 psi g)
4.0 …
400 bar g (58
… 5802 psi g)
400 bar g
(5802 psi g)
600 bar g
(8400 psi g)
400 bar g
(5802 psi g)
400 bar g
(5802 psi g)
600 bar g
(8702 psi g)
Span (continuously
adjustable) or measuring
range, max. operating
pressure (in accordance
with 97/23/EC Pressure
Equipment Directive) and
max. test pressure (in
accordance with
DIN 16086) (max. 120 bar
for oxygen measurement)
Depending on the process connection, the
span may differ from these values
Depending on the process connection, the
measuring range may differ from these values