Cable specifications
● Only use cables with at least the same degree of protection as the sensor to install the
sensor.
● The line length from the cable gland to the terminals must be kept as short as possible. Line
loops in the terminal box must be avoided.
● To guarantee the IP67 degree of protection, use cables with the required specifications.
WARNING
Protective conductor terminal
The required cable is min. AGW16 or 1.5 mm
2
Cu.
WARNING
Wire insulation
For field wiring installation: Ensure that the national requirements of the country in which
the flowmeters are installed is met.
See also
5.1
Remote installation
Note
Remote installation only
The following applies to remote installation of MAG 5000 / 6000 or MAG 6000 I.
Connecting
5.1 Remote installation
SITRANS FM MAG 5100 W
34
Operating Instructions, 01/2020, A5E03063678-AC