Miscellaneous
8 - 3
Siemens-Elema AB
SPR8-215.898.02
Page 3 of 6
MOBILETT Plus/ Plus E
Solna, Sweden
Rev. 01
01.02
SPS-UD
Replacement of mains cable/cable winder
8
•
Remove side covers and the cover with the cassette compartment.
•
Remove the four (two on each side) screws that fixes the cable winder assembly
(pos. 1 / Fig. 2).
•
Pull out the cable winder assembly approximately 10 cm.
•
Disconnect main switch plug K52 (pos. 1 / Fig. 3).
Fig. 2
Fig. 3
Sharp edge!
The brake handle could be damaged.
Lock the brake handle, otherwise it is damaged by the sharp
upper edge of the cable winder assembly.
MOB00618
1
MOB00619
1
CAUTION