
Mechanical systems
10 - 27
Siemens-Elema AB
SPR8-215.898.02
Page 27 of 28
MOBILETT Plus/ Plus E
Solna, Sweden
Rev. 01
01.02
SPS-UD
•
Install the new friction linings according to (Fig. 29 or Fig. 30). There are two different
vertical friction linings. Both are equally suitable for the unit.
•
Reinstall the upper counter bearing.
•
Tighten the screws.
•
Check that there is no braking action on the counter bearings.
•
Tighten the adjustment screw. Make sure that the spring mechanism is under tension
before releasing the ceiling support.
•
Attach the column cover and side covers and the cover with the cassette compartment.
•
Check parallelism, refer to section "Parallelism".
•
Adjust counterbalance, refer to section "Counterbalance".
•
Adjust friction linings, refer to section "Adjusting the friction linings".
Fig. 29
Fig. 30