Service Manual
Maintenance Instructions
Siemens Ltd. Med India
Version 6.0
Copyright
SIEMENS LTD. All rights reserved. For internal use only
8-9
8.10.2 Removing the Front cover.
Turn off the Multimobil 2.5,disconnect the mains chord
and wind the cord around the cable holder.
Lower the X-ray source assembly to transport position.
Lock the Multimobil 2.5 by applying the brake.
Arrange some kind of soft bedding to place the front
cover on .
Unscrew the four ornamental screws with 2.5mm allen
key.
Grasp the cassette box and slowly pull it outside.
Position the front cover on a soft bedding.
8.10.3 Removing the X-ray source assembly.
Turn off the Multimobil 2.5,disconnect the mains chord and wind the cord around the
cable holder.
Lower the X-ray source assembly to transport position. Lock the movement of the
Multimobil 2.5 by applying the foot brake.