![SICK TRANSIC111LP Operating Instructions Manual Download Page 69](http://html1.mh-extra.com/html/sick/transic111lp/transic111lp_operating-instructions-manual_1264716069.webp)
69
8014084/ZWT7/V3-1/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP
Subject to change without notice
ADJUSTMENT
6
6.2
Calibration
The analog output can be frozen for calibration. Use function
Cal.C,
with the keypad. Use the command
Adjust,
see “Hold outputs for calibration (com-
, for input via the serial interface.
6.2.1
Using ambient air
●
Normal ambient air provides a convenient way to calibrate the TRANSIC111LP because
the oxygen concentration of dry ambient air is constant at 20.95 vol% O
2
.
▸
Ensure the sensor is completely in the ambient air. Important: Pay attention to
measured oxygen value of 21.0% O
2
±0.2% O
2
.
▸
Correct the humidity.
shows the calibration display expected for the ambient air as function of the
temperature (ºC) and relative humidity (% r.h.)..
The following Table shows the measured oxygen values (in % O
2
) for a gas concentration of
20.95% O
2
with varying humidity values. The Table shows examples for measured values
when measuring wet gases without r.h. corrections having been entered in the
TRANSIC111LP measuring device (i.e., the relative humidity is set to 0% r.h.). Both gas
dilution effect and r.h. dependency effect are included in the Table.
(% r.h.)
Temp (ºC)
0
10
20
30
40
50
60
70
80
90
100
0
21.0
21.0
21.0
21.0
20.9
20.9
20.9
20.9
20.9
20.9
20.8
5
21.0
21.0
21.0
20.9
20.9
20.9
20.9
20.8
20.8
20.8
20.8
10
21.0
21.0
20.9
20.9
20.9
20.8
20.8
20.8
20.7
20.7
20.7
15
21.0
21.0
20.9
20.9
20.8
20.8
20.7
20.7
20.6
20.6
20.6
20
21.0
20.9
20.9
20.8
20.8
20.7
20.6
20.6
20.5
20.4
20.4
25
21.0
20.9
20.8
20.8
20.7
20.6
20.5
20.4
20.3
20.3
20.2
30
21.0
20.9
20.8
20.7
20.6
20.4
20.3
20.2
20.1
20.0
19.9
35
21.0
20.9
20.7
20.6
20.4
20.3
20.1
20.0
19.8
19.7
19.6
40
21.0
20.8
20.6
20.4
20.2
20.1
19.9
19.7
19.5
19.3
19.1
45
21.0
20.8
20.5
20.3
20.0
19.8
19.5
19.3
19.1
18.8
18.6
50
21.0
20.7
20.4
20.1
19.7
19.4
19.1
18.8
18.5
18.2
17.9
55
21.0
20.6
20.2
19.8
19.4
19.0
18.6
18.3
17.9
17.5
17.2
60
21.0
20.5
20.0
19.5
19.0
18.5
18.1
17.6
17.1
16.7
16.2
65
21.0
20.4
19.7
19.1
18.5
17.9
17.3
16.8
16.2
15.6
15.1
70
21.0
20.2
19.4
18.7
17.9
17.2
16.5
15.8
15.1
14.4
13.8
75
21.0
20.0
19.1
18.2
17.3
16.4
15.5
14.7
13.8
13.0
12.2
80
21.0
19.8
18.7
17.5
16.5
15.4
14.4
13.4
12.4
11.4
10.4
Table 11: Oxygen measured values for relative humidity