8014084/ZWT7/V3-1/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP
Subject to change without notice
4
OPERATION
4.7
Output of measuring results
4.7.1
Start continuous output (command R)
Starts RUN mode. Outputs values defined with the command
FORM
ing results (command FORM)”, page 45
INTV
(
see “Display/set continuous out-
put interval (command INTV)”, page 42
) defines the output interval. Command
S
(
continuous output (command S)”, page 42
) stops RUN mode.
Syntax:
R<cr>
Example:
4.7.2
Stop continuous output (command S)
Stops RUN mode and switches the serial output to STOP.
Syntax:
S<cr>
Example:
4.7.3
Display/set continuous output interval (command INTV)
Sets the frequency for measured value output in RUN mode (
Syntax:
INTV [Value] [Unit]<cr>
Example:
4.7.4
Send measuring results (command SEND)
Outputs the last results (in accordance with
FORM
,
see “Format measuring results (com-
) in STOP mode. The command can be used with an address in
POLL mode.
Syntax:
SEND [Address]<cr>
SEND [Formatting string]<cr
Example:
>r
Oxygen = 21.0
Oxygen = 21.0
Oxygen = 21.0
>S
>
Value
=
Time interval in which results are output (0 ... 255)
Unit
=
Interval time unit, S for seconds, MIN for minutes or H for hours
>intv
INTERVAL : 1 ? 5
UNIT S ? min
Address
=
Device address
Formatting string
=
Character string specifying the output format for the measuring results
>send 20.9 20.8 24.5