Instrument Air Conditioning Unit | SICK

8024118/V1-0/2019-03/en

2

6. 

Insert the new filter element (

6

to the marking (

5

) on the filter 

bowl.

7. Grease the sealing rings at the 

filter element (

7

) as shown.

Grea

se

8. 

Check marker at the filter top (

8

).

9. 

Grease the new filter bowl o-ring 

on the filter top (

9

) and attach 

the UFM-D05 housing.

10. 

Check that the final position of 

the filter housing is locked (

ß

).

Gr

ea

se

7.3 Replacing the UFM-D05 maintenance kit

1. Remove UFM-D05 (

2

) by pressing 

the arresting hook (

1

).

2. Unfasten UFM-D05 from the 

outlet (

5

) and remove the design 

shell (

4

).

3. Detach the maintenance unit (

6

from the tubing at the inlet by 

removing the union nut.

4. Install the new maintenance unit 

(

6

) in reverse order.

7.4 

Replacing the filter element DF-T 0050 OS

1. Loosen and remove lower part 

(bowl) of the filter housing (

1

).

2. 

Remove the filter element (

2

).

3. 

Insert the new filter element (

4

to the marking (

3

) on the filter 

bowl.

4. Grease the sealing rings at the 

filter element (

5

) as shown.

Gr

ea

se

8. 

Specifications

Instrument air specifications 

Particle (class 6/ISO 8573:2010)

≤ 100.000 solid particle 

[

1 ... 5 µm

]

Water vapor (class 4/ ISO 8573:2010)

≤ +3 °C pressure dew point (steam)

Total oil portion (class X/ ISO 8573:2010)

≤ 10 mg total oil (liquid, aerosol, fog)

Flow rate 

1.3 Nm

3

/h

Pressure

12.5 bar

Operating temperature

1 ... 60 °C

Volume

1.5 l

8.1 Dimensional drawing

196

 (7.72)

465

 (18.31)

9

 (0.35)

7

 (0.28)

543

 (21.38)

400

 (15.75)

151

 (5.94)

125

 (4.92)

125

 (4.92)

125

 (4.92)

6.  Commissioning

 

b

Before connecting the instrument air, blow the instrument air 

into the environment for a few minutes. Check dryness with a 

paper towel. 

 

b

Slowly apply pressure to the instrument air inlet. For the maxi

-

mum limit value refer to “Specifications”.

 

b

Make sure that no pressure peaks occur. See manufacturer’s 

documents attached.

7.  Maintenance

7.1 Yearly maintenance work

Only the replacement of the filter elements of the maintenance 

kit has to be carried out once a year. 

Requirements:

 maintenance kit, part No. 5339994.

The maintenance kit consists of the components: 

 

• Filter element DFM-S 0070 and housing sealing

 

• Maintenance kit for condensate separator UFM-D05 and 

sealing

 

• Filter element DF-T 0050 and housing sealing

Overview of maintenance kits elements

1. 

Condensate separator UFM-D05

2. 

Filter DFM-S 0070 OS

3. 

Filter element DFM-S 0070 OS

4. 

Filter element DF-T 0050 OS

5. 

Filter DF-T 0050 OS

Fig. 3: Disassembly/assembly of maintenance kits components

  Warning: Risk of death by electric shock

Risk of electric shock when working on electrical parts while 

the unit is switched on. 

 

b

Before any action step ensure that the unit is turned off. 

  Caution: Risk of injury due to high pressure

When working on parts under pressure, it is possible that 

parts come off unexpectedly or compressed air escapes.

 

b

Before working on the unit, ensure an appropriate decom-

pression within the instrument air tubes. 

7.2 

Replacing the filter element DFM-S 0070 OU

1. Slowly reduce and stop the pressure supply at the instrument 

air inlet.

2. Remove the UFM-D05 (

2

) by 

pressing the arresting hook (

1

).

3. Reset UFM-D05: press the TEST 

button (

3

) below the LED, hold it 

for at least 5 sec. and switch off 

the voltage supply.

4. Loosen and remove the lower part 

(bowl) of the filter housing (

3

).

5. 

Remove the filter element (

4

).