DECOMMISSIONING
9.1.3
Switching off
▸
Switch off the control cabinet cooling unit.
▸
Switch off the PowerCEMS100 analysis system at the main switch (on the outside of the
door) (position “OFF”) and secure against switching on.
9.1.4
Decommissioning
▸
Disconnect the PowerCEMS100 analysis system completely from the power supply.
▸
Disconnect the sample gas lines from probe and system cabinet.
▸
Dispose of the condensate.
▸
Empty the humidifier containers.
▸
Close off the gas inlet/outlet on the analyzer.
▸
Close off the end of the sample gas line on the probe.
▸
Dismantle the sample gas probes and close off the connections with blind flanges.
9.1.5
Storage
Storage conditions
●
In dry, vented rooms.
●
Ambient temperatures: -20 ... +55 °C
●
Relative humidity:
≤
75% without condensate.
Note
▸
Observe the minimum switch-off times for the control cabinet cooling unit.
▸
Non-observance of the minimum switch-off times can damage the control
cabinet cooling unit.
Refer to the relevant Operating Instructions for information on minimum switch-
off times of the control cabinet cooling unit.
DANGER
SAMPLE GAS DETRIMENTAL TO HEALTH OR TOXIC
Serious injuries of the respiratory system or death.
▸
Wear suitable respiratory protection for all work on the gas lines.
▸
Prior to decommissioning, adequately purge the measuring system
with neutral gas (e.g. nitrogen N
2
)
WARNING
ACIDIC CONDENSATE
Serious injuries and/or serious damage to environment and nature.
▸
Wear personal protective equipment (gloves, respiratory protection).
▸
Disposal according to local and legal environment regulations and
guidelines.
Note
Prior to storing the PowerCEMS100 analysis system, purge the sample gas
paths in the analysis system with a “dry”, neutral gas.
This helps to prevent condensation.