5.4.3
Adjustment/Automatic
Menu 2.3: Adjustment/Automatic
Automatic adjustments are started in this menu item (parameter settings
SOPAS ET).
Description of programs
No.
Name
Function
1
Optical Adjustm.
Adjustment of zero point (with zero gas) and span point (with internal adjust-
ment cell)
2
Zero: Analyzer
Check of zero point with internal zero point reflector.
3
Zero: System
Check of zero point by zero gas feeding with/without gas sampling system
(as configured in SOPAS ET).
4
Adjust. Analyzer
Check of span point with internal adjustment cell.
5
Adjust. System
Check of span point by feeding test gas.
6
Drift Check
Drift check without acceptance of values (with internal adjustment cell or test
gas generator (preset)).
7
Drift Correction
Drift check with automatic acceptance of correction (with internal adjust-
ment cell or test gas generator (preset)).
8
Cancel
Cancel test or adjustment.
2.3
Hg
5,5 μg/m3
34
1
To start the check: Select the program and press “
Start
”.
2
The password (
p. 36, §4.6) is queried.
3
Operating state: “
Maintenance
”.
4
The Measuring screen is displayed during the check (with a down
counter to the end of the check).
5
After the end of the adjustment, the device switches again to oper-
ating state “Measuring” (if “
Maintenance
” was previously manually
set: Again to “
Maintenance
”).
●
If the deviation exceeds a limit (parameter settings in SOPAS ET),
the device switches to the “
Maintenance request
“ classification
Measuring
Back
Start
Menu
Automatic
/Adj/Autom
1 Optical Adjustm.
2 Zero: Analyzer
3 Zero: System
4 Adjust. Analyzer
5 Adjust. System
6 Drift Check
7 Drift Correction
8 Cancel
Optical Adjustm.
●
Measuring screen with down counter to the end of the program [s].