Product Description
MERCEM300Z · Operating Instructions · 8013255 V2.0 · © SICK AG
17
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
2.2.1
Method of operation
The device operates independently.
The operating states are signaled by status signals and entered in a logbook (SOPAS ET
●
Feeding of the sample gas to the analyzer in a heated sample gas line.
●
Conversion of mercury compounds into elementary mercury in a thermal converter.
●
Measuring principle: Photometric with Zeeman effect.
●
Standardized (pressure, temperature) output of measured values.
●
Output, relative to humid flue gas.
●
Test gas feed (adjustment)
●
Integrated adjustment cell for automatic adjustment without test gases (
).
●
Operation with external UPS (uninterruptible power supply) (
2.2.2
Test gas generator CALSIC300 (option)
Option: Integrated test gas generator CALSIC300 (position
●
The initial solution, an HgCl
2
solution, is in a storage container (position
).
●
Using a vaporizer, the test gas generator generates test gas from the HgCl
2
solution.
●
The test gas is transported to the cell via the gas sampling system (“test gas feeding on
gas sampling system”) (
●
The test gas concentrations of the test gas generator are set using a menu.
●
Test gas feeding is started cyclically or manually (
2.2.3
Instrument air conditioning (option)
Option: Integrated instrument air conditioning.
Menu item
Test gas feed ...
Reference
Test
... manually with internal test gas generator CALSIC300
(option)
manual
... manually with external test gas feed on gas sampling
system
p. 26, §3.5.3 and
automatic
... cyclically with internal test gas generator CALSIC300
(option)
For more information on instrument air conditioning
Operating Instructions
for instrument air conditioning.