A D D E N D U M   T O   O P E R A T I N G   I N S T R U C T I O N S

MARSIC300

Ship Emission Measuring Device

Maintenance

Summary of Contents for MARSIC300

Page 1: ...A D D E N D U M T O O P E R A T I N G I N S T R U C T I O N S MARSIC300 Ship Emission Measuring Device Maintenance...

Page 2: ...the limits of the legal determination of Copyright Law Any modi fication abridgment or translation of this document is prohibited without the express written permission of SICK AG The trademarks state...

Page 3: ...Replacing the Electronics module 12 5 Consumable wearing and spare parts 13 5 1 Housing 13 5 2 Analyzer 14 5 3 Gas sampling unit SFU 15 5 4 Accessories for external instrument air conditioning FRL SF0...

Page 4: ...est gases The technician must be able to avoid hazards caused by noxious test gases The technician must be familiar with gas lines PTFE lines and their screw fittings be able to ensure gas tight conne...

Page 5: ...rsonal injury or death CAUTION Hazard or unsafe practice which could result in less severe or minor injuries Notice Hazard which could result in property damage Note Hints 1 4 3 Information symbols Ta...

Page 6: ...ions have been read and understood Contact SICK Service if you have any questions b Refer to the device s Declaration of Conformity for the applicable standards and directives b Keep this document tog...

Page 7: ...ee Operating Instructions Section 8 6 1 M Analyzer cabinet Check the filter pads Clean or replace the filter pads as necessary See Operating Instructions Section 8 5 Check the exhaust hose to see if b...

Page 8: ...See Operating Instructions Section 7 4 6 Complete Service Report and Checklist 2 Y2 3 Analyzer cabinet Replace the filter pads See Operating Instructions Section 8 5 Measuring cell Replace the cell i...

Page 9: ...ve the water trap from the bracket by pressing the bracket down onto the connection 6 Note the date of installation on the new water trap 7 Insert the new water trap together with the bracket into the...

Page 10: ...Swap the register when the values were not transferred correctly In case of incorrect transfer swap the register 1 Menu Parameterization Select Modbus 2 Register Swap Use the drop down list to change...

Page 11: ...hen the device is switched off Prerequisites The device is switched off Procedure 1 Pull cover 1 off 2 Replace the filter pad inside 3 6 Reference energy NOTICE Reference energy must not be reset with...

Page 12: ...be used when the firmware versions of the old and new device are identical Procedure b See Operating Instructions Section 9 14 4 TROUBLESHOOTING 12 A D D E N D U M T O O P E R A T I N G I N S T R U C...

Page 13: ...Description Part No Display module complete 2081834 Filter fan 230 V 6029937 Filter pad 173 mm x 173 mm 5306678 Cabinet outlet filter 204 mm x 204 mm x 30 mm 5315501 CONSUMABLE WEARING AND SPARE PART...

Page 14: ...5339 2097393 2081463 2081836 5310504 5329115 6058571 2109587 Figure 3 Overview interior 5 CONSUMABLE WEARING AND SPARE PARTS 14 A D D E N D U M T O O P E R A T I N G I N S T R U C T I O N S MARSIC300...

Page 15: ...s valve 2109587 2 2 way solenoid valve for test gas 6058571 Maintenance kit Cell input filter 2045339 Non return valve cell filter incl nozzle and spring washer 2097393 Hose kit material PTFE incl fit...

Page 16: ...ated coal element for instrument air conditioning 5320896 5341738 Activated carbon filter complete with sight glass for instrument air conditioning 5320896 5342518 Replacement filter glass with O ring...

Page 17: ...s line 5 m 115 V 6035905 Heated sample gas line 10 m 115 V 6035910 Heated sample gas line 15 m 115 V 6035915 Heated sample gas line 20 m 115 V 6035920 Heated sample gas line 25 m 115 V 6035925 Heated...

Page 18: ...Cable ties 2081755 Hose kit material PTFE incl fitting material 2 m hose outer 6 mm inner 4 mm material PTFE 5 m hose outer 10 mm inner 8 mm material PTFE Hose cutter 2 x cap nuts clamping rings supp...

Page 19: ...tion Supply MAINS UPS 3 pole Power supply for electronics unit internal MAINS 5 pole External power supply F1 Internal F2 Internal More information Connections and interfaces of further modules b See...

Page 20: ...ead 3 6 4 6 A2 50 A4 50 5 6 Alu A2 70 A4 70 A2 80 A4 80 8 8 Titan 10 9 12 9 M 1 6 0 4 0 05 0 05 0 05 0 11 0 16 0 19 0 26 0 31 M 2 0 45 0 1 0 1 0 11 0 22 0 32 0 39 0 55 0 66 M 2 5 0 45 0 21 0 22 0 23 0...

Page 21: ...tory setting Option Freeze measured values 7 1 1 Adjustment procedure The bellows valve on the gas sampling unit SFU is closed The maintenance signal is activated The measuring system is purged with i...

Page 22: ...matically jumps to 999 as soon as the CO2 value is 1 Vol The SO2 CO2 ratio returns to normal after the adjustment is completed The mainte nance signal goes out and measuring operation can be continued...

Page 23: ...completed until the measured values are acti vated The maintenance signal goes out and measuring operation can be continued when the initial state is reached Measuring Mode Premeasuring Mode Zero Poin...