4 Set up with the display and adjustment module
LFR • Two-wire 4 … 20 mA/HART
62270-EN-191108
Menu item
Parameter
Default setting
Current output
Current output -
Mode
Output characteristics
4 … 20 mA
Reaction when malfunctions occur
≤ 3.6 mA
Current output -
Min./Max.
3.8 mA
20.5 mA
Lock adjustment
Released
Menu item
Default setting
Menu language
Order-specific
Displayed value 1
Filling height in %
Displayed value 2
Electronics temperature in °C
Backlight
Switched on
Menu item
Parameter
Default setting
Sensor status
-
Peak value indi-
cator
Distance
-
Measurement re-
liability
-
Peak values, ad-
ditional
Temperature
-
Curve display
Echo curve
-
False signal sup-
pression
-
Simulation
Percent
Echo curve mem-
ory
Percent
Menu item
Parameter
Default setting
Date/Time
Actual date/Actual time
Reset
-
Copy instru-
ment settings
-
Scaling
Scaling size
Volume in l
Scaling format
0 % corresponds to 0 l
100 % corresponds to 0 l
Current out-
put 1
Current output - Meas. vari-
able
Lin. percent - Level
Current output - Adjustment
0 … 100 % correspond to
4 … 20 mA
Display
Diagnostics
Additional adjustments