Gas connections
Protective gas connections
Sample gas connections
Approval conditions for sample gas feed
Approval conditions for purge gas feed for an Analyzer module
6.6
Mains connection
Connection
Designation
Suitable for
Inlet:
Swagelok 8 mm
Metal tube with 8 mm outer Ø
Outlet for Enclosure GMS815P-PS-2G:
Thread G 1"
Screw fitting
Outlet for Enclosure GMS815P-PS-3G:
Thread G
3
/
4
"
Screw fitting
Connection
Designation
Suitable for
Standard:
Swagelok 6 mm
Metal tube with 6 mm outer Ø
Option:
Swagelok ¼“
Metal tube with ¼“ outer Ø
Parameter
Allowable value
Sample gas pressure in enclosure:
–500 … +1000 hPa (–0.5 … +1.0 bar)
Sample gas volume flow:
Max. 100 dm
3
/hour
Parameter
Allowable value
Maximum purge gas pressure in
enclosure:
15 hPa (15 mbar)
●
Protective gas connection position
●
Protective gas feed parameters
●
Sample gas connection positions
●
Sample gas connections function
→
“GMS800 Series” Operating Instruc-
tions
●
Other specifications for sample gas
→
Supplementary Operating Instruc-
tions for the Analyzer module fitted
Mains voltage
– Enclosure GMS815P-PS-2G:
120 V AC ± 10%
or
[1]
230 V AC ± 10%
[1] Depending on device version; refer to type plate for appropriate value (
– Enclosure GMS815P-PS-3G:
115 V AC ± 10%
230 V AC ± 10%
Mains frequency (AC):
47 … 63 Hz
Allowable overvoltages:
Transient overvoltages in supply network must not exceed Over-
voltage Category II according to IEC 60364-4-443
Power input:
50 VA / max. 300 VA
Internal mains fuses
– Primary:
6.3 A (not exchangeable)
[2]
[2] Replace the power supply unit after triggering
– Secondary:
10 A (exchangeable fusible cutout)
[3]
[3] F1 on the “fuse board” – spare part: “ET fuse F10A0“, Part No. 2062251.
Required connection cable
– Conductor cross-section:
≥
0.75 mm
2
– Version:
IEC 60227 or IEC 60245