9.5
Calibration
Menu "Calibration" serves to retrieve the following submenu items:
•
Zero Adjust: Zero adjust
•
SPAN Test: Manual SPAN test
•
EPA SPAN Test: SPAN test according to EPA Guidelines
9.5.1
Performing zero adjust
Submenu item "Zero Adjust" serves to perform the zero adjust.
NOTE
•
The adjust process is started when the prompt "Are you sure to start adjust proce‐
dure" is confirmed with "Enter". Making entries during the process or aborting are
not possible.
•
The adjust first starts when the device temperature stabilizes at 60 °C (± 0.5 °C).
Warming up time depending on ambient conditions; 30 - 60 minutes.
•
Press "Enter" to save the data when the zero adjust has completed.
back
Calibration
back
Start:Enter
back
back
Save:Enter
Zero Adjust
Zero Adjust
Zero Adjust
Zero Adjust
Zero Adjust
Are you sure to
start adjust procedure?
Parameters
back
Zero Adjust
Span Test
EPA Span Test
Caution: Operation
temperature not valid
T: 61.5 °C
Start:Enter
Please wait!
..................
T = 60°C± 0.5 %
Please wait
Amplifier Values
Amp1: 0 Amp2: 6
..................
C1 : +0,0...
C1 -var : +0,0...
C2 : +0,0...
Figure 56: Performing zero adjust
9
MENUS
70
O P E R A T I N G I N S T R U C T I O N S | GM901-02