Checking status LEDs, error and warning messages
NOTE
Quick check of Measuring screens on the LC display of the evaluation unit:
b
Using the Operating Instructions,
see "Operating and display elements (evaluation
.
When error or warning messages are present:
b
Determine and clear the error cause using the Operating Instructions,
interface commands - Diag", page 65
, and the Error and Warning Message
Tables,
and
see "Warning messages", page 84
If the error cause cannot be cleared:
b
Contact SICK Service.
7.11.1
Operating states
Message on EvU display
Meaning
INIT
Initializing the evaluation unit
INITIALISATION
Initializing the SR-unit
SIGNAL_ADJUST
Adjusting amplification to changed transmission
MEASURING
Measuring operation
MEASURING LL
Measuring operation with active tracking of the laser
beam length (line locking)
MEASURING LD
Measuring operation; beam length tracked (line locking
done)
MAINTENANCE
Maintenance mode (no measurement)
DOWNLOAD
Download new software (for Service purposes only)
7.11.2
Select the ambient temperature range
NOTICE
Temperature stabilization: 30 minutes
The device stabilizes the temperature after the temperature is modified.
•
Warning message “DEV TEMP” is output during temperature stabilization.
1
Activate Parameter mode.
2
Menu path: Setting -> Ambient Temp.
3
Select the temperature range from the Table that matches the ambient conditions
best at the measuring location,
see "Possible ambient temperatures", page 99
.
COMMISSIONING
7
8019532/100Q/V1-1/2018-08 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM700 Ex
57
Subject to change without notice