8.4.4
Serial interface commands - Diag
The serial interface is only used when fast access to the menu in the EvU is not possi‐
ble due to Ex protection. The following Chapters explain a selection of commands the
user will need to analyze a problem when errors occur.
8.4.4.1
Table with warning messages
Call up the Table of warning messages.
Syntax:
GM700<<DIS WTBL
8.4.4.2
Call up the Table with error messages
Call up the Table with error messages
Syntax:
GM700<<DIS MTBL
8.4.4.3
View the device configuration
Call up the list of the device configuration
Syntax:
GM700<<DIS CONFIG
8.4.4.4
Check the system state and configuration
Output of the current system state and configuration
Syntax:
DIS DEV PARA
8.4.4.5
Call up diagnostic data
Command “DIS MERB” calls up the diagnostic data.
Syntax:
GM700<<DIS MERB
8.5
Setting the display contrast
The rotary potentiometer for setting the display contrast is located above the “Digital
Out” terminal block.
b
Use a 2 mm precision screwdriver to adjust the display contrast setting.