A D D E N D U M T O O P E R A T I N G I N S T R U C T I O N S | FWE200DH
Subject to change without notice
3
COMMISSIONING / CONFIGURING
3
Commissioning / configuring
3.1
Temperature setting of the heating bands
The heating band settings (
) must first be made in the software
settings. The application-related settings must then be made in the control unit. For this
purpose, it is necessary to locate the heating band setting in the control unit and make the
setting with a slotted screwdriver.
Fig. 7: Heating band settings, SOPAS ET (Parameters / System Parameters)
Fig. 8: Heating band settings on control unit
WARNING: Hot gases may escape
Risk of burns from escaping hot gases.
▸
Open inlets of gas-carrying parts must be closed during device installation, assembly
and commissioning to prevent leakage.
▸
Suitable protective measures against any escaping sample gas must be taken.
▸
Personal protective equipment must be worn.