4.1
Operation and Menu structure of the SPU via LC-Display
4.2
Operational concept
FLOWSIC60 is equipped with a front panel containing a two-line LCD to display current
measured values and diagnostics and enter installation specific parameters. The values to
be displayed or modified can be selected using the magnetic pen without removal of the
front cover.
Fig. 21
Front panel
The sensors for operation with the magnetic pen are highly sensitive. In order to navigate
accurately through the menus, consider the instructions below:
Hold the magnetic pen right-angled to the front panel.
Point the magnetic pen tip directly onto the sensors. (not the keypad)
Remove the magnetic pen at least 10 cm from the display after each operation of a
sensor.
Hold magnetic pen onto a sensor for approx. 1 s means short operation of sensors.
Hold magnetic pen onto a sensor for > 2s means long operation of sensors.
Operation of FLOWSIC60 using the LC-Display is realized by four commands.
Refer to
for description of these commands.
LC-Display
Sensors for operation with magnetic pen
Table 1
Display commands
Sensor
Command
Function
Description
Step (Short)
STEP +
Step forward
Use the STEP + command to navigate forward
through the menu
Data (Short)
STEP -
Step backward
Use the STEP - command to navigate backward
through the menu
Data (Long)
ENTER
Enter
Use the ENTER command to switch FLOWSIC60
into Configuration mode, select a menu level or
edit any modifiable parameter.
C/CE (Short) RETURN
Cancel/Menu level up
Use the RETURN command to cancel parameter
editing or to return from any menu to the upper
level; continue pressing to return to the initial
display