4.2.4
Meter installation
Depending on the meter orientation and the upstream piping, a distinct setup code has to
be set. An overview of all options is provided in
.
For instance, for a gas meter being vertically installed in straight upstream piping, the set-
).
The setup code is 3 for a gas meter with horizontal orientation and elbow from the top. The
default setup code is 0.
Table 5
Setup code depening on meter orientation and upstream piping
In order to change the parameter for the meter installation, proceed as follows:
1
“1 Device setup” > “3 Setup” > “7 Volume Flow” > “3 Setup Code” > Enter Setup Code
2
Press ENTER
3
Press SEND to write the new setup code to the meter
4.2.5
Pressure value
In case that no pressure sensor is available, the device can be switched to a mode with
fixed pressure value:
1
“1 Device setup” > “3 Setup” > “8 Pressure” > “1 Pressure Source” > Select “Fix Value”
2
Press ENTER
3
Press SEND to write to the meter
4
Select in the same menu “8 Pres Fix/Repl. Val.”
5
Insert the fixed pressure value to be used
6
Press ENTER
7
Press SEND to write to the meter
4.2.6
HART
®
dynamic variables
In order to change the HART
®
dynamic variables (PV, SV, TV, QV) to your needs, follow below
steps:
1
“1 Device setup” > “3 Setup” > “4 HART/Loop current” > “Dyn. Var. Assign.”
2
Chose the corresponding variable (e.g. 1 PV) and select from the list
3
Press ENTER
4
Press SEND to write the value to the meter
Configuration of upstream
piping
Meter orientation
Horizontal
Vertical
Straight (>20D)
0
5
Elbow right
1
6
Elbow left
2
7
Elbow from top
3
8
Elbow from bottom
4
9