Commissioning and parameterization
4.3.1.2
Digital outputs
Select the ”Configuration / I/O Configuration / Digital Output / Digital outputs” directory for
entering parameters.
Fig. 77
”Configuration / I/O Configuration / Digital Output / Digital outputs” directory
Field
Parameter
Remark
Inverted
Inactive
Define the switching direction
Active
Source bit
Bit 0
Malfunction
Bit 1
Maintenance
Bit 2
Maintenance request
Bit 3
Function check
Bit 7
Operation ( no malfunction)
Bit 16 to 31
Aim bit of the limit value switch (
)
Source sensor
Selection of the component:
- Sensor 1 to 8 when the device status is to be output
- MCU when limit values are to be signaled