78
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Start-up
8.1
Safety information for start-up
8.2
Preparing for start-up
Start the peripheral devices (e.g. heated sample gas line, registration devices). Wait
until they are ready for operation as necessary (e.g. warm-up time).
Check availability for the auxiliary gases (pressure, reserves).
Activate auxiliary gas supply (e.g. open valves).
Start the H
2
leakage monitoring device and check the function.
Close all housings.
8.3
Activation procedure
To start operation of the EuroFID:
Switch on the external mains switch or the mains
switch on the rear of the operating unit.
Automatic procedures after activation
➝
The electronic circuit in the analyzer unit starts the measuring system.
➝
The EuroFID software starts in the operating unit and runs a self-test.
➝
Information on the installed software version is then displayed:
➝
When the software has started, the “System in Service” status is activated (+ LED
“Service” lights up).
➝
The sensor block in the analyzer unit starts to heat up.
➝
The supply of instrument air is activated in the analyzer unit (internal solenoid valve
opens).
➝
The supply of zero gas is activated in the analyzer unit.
CAUTION:
Possible danger for connected devices/stations
When the operating unit is switched on, the analog outputs emit a current of
22 mA for a short time (= 110 % of the measuring range).
Before switching the operating unit on:
If necessary for safety reasons,
inform connected stations and/or deactivate connected devices (e.g. alarm
indicator).
E U R O F I D
S o f t w a r e R e v . 5 . 2
C o n f i g u r a t i o n 5 . 2
V 5 . 2 0 2 . 0 7 . 0 3
T y p e e m i s s i o n s c n t r l
5 5 6 2
P r o d . d a t e 1 1 . 2 0 0 6
Device name
Software version
1
Software version / Release date (Day.Month.Year)
Software type
Date of manufacture of the operating unit (Month.Year)
1
EU type approved EuroFID versions have the software version 4.12084 ( EU type approved software version).
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...