50
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Installation of the Analyzer Unit
4.4.4
Installing the analyzer
1
Check whether the sealing sleeve (silicone foam seal) is mounted on the sample gas
probe (
).
2
Push the analyzer into the housing (place on the rails); carefully insert the sample gas
probe through the flange opening.
3
Connect the gas supply lines internally: Connect the three couplings of the gas lines in
accordance with the color code of the hoses (red/blue/white).
4
Insert the exhaust gas line (
recommendation:
PTFE tube) from the outside through the
plastic bulkhead fitting into the industrial housing; place the clamping ring of the plastic
bulkhead fitting onto the exhaust gas line.
5
Connect the exhaust gas line to the “Exhaust” gas connection of the analyzer.
6
Push the analyzer into the end stop.
7
Place the clamping bar onto the rails behind the base frame.
8
Press the clamping bar and the analyzer firmly against the flange. In this condition
secure the clamping bar in place (tighten the clamping screws).
9
Connect the connection cable of the analyzer to the corresponding counterpart on the
inside of the industrial housing.
10
Close the plastic bulkhead fitting around the exhaust gas line.
Fig. 13
Installing the analyzer
1
2
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...