PROJECT PLANNING
4
Project planning
4.1
Device configuration
The device components required for a measuring system depend on the respective
application conditions. The following Section can help you with your selection.
4.1.1
Selecting the sender/receiver unit
Selection of the suitable sender/receiver unit depends on the wall and insulation thickness
of the duct and the composition and temperature of the sample gas.
Table 7: Sender/receiver unit selection Table
Wall and
insulation
thickness
[mm]
Nominal
length NL
[mm]
Protective
tube
length
[mm]
Exhaust gas, exhaust air
Sender/receiver unit
type
Max.temperature
in °C
Composition
Max. 100
435
300
220
Non-corrosive/
low corrosion
DHSP-T2V11EX2K
Corrosive
DHSP-T2H11EX2K
400
Non-corrosive/
low corrosion
DHSP-T4V11EX2K
Max. 400
735
600
220
Non-corrosive/
low corrosion
DHSP-T2V21EX2K
400
Non-corrosive/
low corrosion
DHSP-T4V21EX2K
Further variants available on request
●
Select the nominal length of the sender/receiver unit so that the measuring opening
is an adequate distance away from the internal duct wall.
– Standard version clearance: At least 100 mm.
– High-temperature version clearance: Between 100 mm and 140 mm so that the
measuring device does not protrude too far into the duct and is not warmed up
unnecessarily by the hot sample gas.
●
The selected nominal length of the sender/receiver unit should only be as long as
necessary, the measuring opening (
see “DHSP-T2VxxEX2K sender/receiver unit”,
) need not be in the middle of the duct.
●
Observe the maximum permissible process pressure (
), safe operation of the measuring system is not possible when this value is
exceeded.
●
Limit values for corrosive gas composition (standard values assuming normal
pressure and temperature conditions, use lower values for mixtures of several
components):
– HCl:
10 mg/m³
– SO
2
:
800 mg/m³
– SO
3
:
300 mg/m³
– NOx:
1000 mg/m³
– HF:
10 mg/m³