Function Commands
<STX><0x03><0x32><ETX>
Laser On (Sleep mode)
<STX><0x03><0x33><ETX>
Laser Off (Sleep mode)
<STX><0x03><0x35><ETX>
Preset activation
9.5.3
Command examples (Standard protocol)
Value binary request
to DME
<STX><0x01><0x21><ETX>
from DME
<STX><0x81><0x21><0x..>… <0x..><0x..><0x..><ETX>
example terminal
5378,8 mm
⇒
81210000D21C
(14 bytes: 02 38 31 32 31 30 30 30 30 44 32 31 43 03)
Value BCD request
from DME
<STX><0x01><0x22><ETX>
to DME
<STX><0x81><0x22><0x..>… <0x..><0x..><0x..><ETX>
example
5378,8 mm
⇒
8122+0053788
Value binary continuous on
from DME
<STX><0x05><0x21><0x01><ETX>
to DME
<STX><0x85><0x21><0x01><ETX>
after that from DME
continuous
<STX><0x03><0x21><0x..>… <0x..><0x..><0x..><ETX>
Value binary continuous off
to DME
<STX><0x05><0x21><0x00><ETX>
from DME
<STX><0x85><0x21><0x00><ETX>
BCD continuous on
to DME
<STX><0x05><0x22><0x01><ETX>
from DME
<STX><0x85><0x22><0x01><ETX>
after that from DME
continuous
<STX><0x03><0x22><0x..>… <0x..><0x..><0x..><ETX>
Value BCD continuous off
to DME
<STX><0x05><0x22><0x00><ETX>
from DME
<STX><0x85><0x22><0x00><ETX>
Temperature
to DME
<STX><0x01><0x26><ETX>
from DME
<STX><0x81><0x26><0x..><ETX>
example
internal tempe54
°
C
⇒
812636
⇒
0x36 = 54
Reception level
to DME
<STX><0x01><0x23><ETX>
from DME
<STX><0x81><0x23><0x..><ETX>
example
-39 dB
⇒
8123D9
⇒
0x100
−
0xD9 = 0x27 =39
Service-status
to DME
<STX><0x01><0x23><ETX>
from DME
<STX><0x81><0x23><0x..><ETX>
example
Icon MF2 and icon PLB is visible on the display:
812542
⇒
0x42 = 10000010 binär
MF2
MF1
TMP
LSR
BUS
ATT
PLB
RDY
1
0
0
0
0
0
1
0