Technical Data
DEFOR · Supplementary Operating Instructions · 8013029 V2.0 · © SICK AG
25
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
5.3
Technical gas specifications
5.3.1
Sample gas
5.3.2
Purge gas
Only valid for version with process cuvette (
)
5.4
Materials with sample gas contact
Allowable sample gas temperature:
[1]
[1] On the sample gas inlet. Keep constant during operation.
– Minimum:
5 °C (41 °F)
[2]
[2] When a sample gas cooler is used: Always above the cooler temperature (dew point).
– Maximum:
55 °C (131 °F)
[3]
[3] With “heated sample gas path” option: Up to +80 °C, depending on the temperature set.
Allowable sample gas dew point:
Below ambient temperature
Particles in the sample gas:
Sample gas should be free from dust and aerosols
[4]
[4] On the sample gas inlet.
Allowable sample gas pressure
[5]
[5] Relative to the ambient/atmospheric air pressure.
–200 … +300 hPa (–0.2 … +0.3 bar)
Sample gas volume flow
– Minimum:
20 l/h (333 cm
3
/min)
– Maximum:
120 l/h (2000 cm
3
/min)
– Recommended:
30 … 60 l/h (500 … 1000 cm
3
/min)
– Standard:
30 l/h (500 cm
3
/min)
Suitable purge gas:
Dry inert gas (chemically neutral gas/gas mixture
without condensable components)
Allowable purge gas pressure
[1]
[1] Relative to the ambient/atmospheric air pressure.
15 … _30 hPa
Purge gas volume flow
– Minimum:
20 l/h (333 cm
3
/min)
– Maximum:
100 l/h (167 cm
3
/min)
– Recommended:
20 … 60 l/h (333 … 1000 cm
3
/min)
– Standard:
30 l/h (500 cm
3
/min)
Component
Material
Cuvette:
Aluminium or stainless steel
[1]
[1] Depending on version.
Optical window:
CaF
2
or quartz
Synthetics:
Viton B, PVDF, PTFE