8.2
Completely replacing the safety laser scanner
1.
Disconnect the connecting cables the system plug.
2.
Unscrew the fixing screws and remove the defective safety laser scanner.
3.
Mount the new safety laser scanner, see “
Mounting the safety laser scanner
”.
4.
Reconnect the connecting cables to the system plug.
5.
Configure the safety laser scanner.
6.
Perform commissioning again, taking particular care to conduct all of the thorough
checks described. For more information, see the operating instructions.
9
Replacing the system plug
NOTICE
Enclosure rating IP65 only applies if the safety laser scanner is closed and the system
plug is mounted.
b
Mount the system plug and the cover plate.
b
Close each M12, RJ45 and SCRJ plug connector on the safety laser scanner using
an IP65 rated male cable connector or a protective cap.
°
Tightening torque for M12 plug connector: 0.4 Nm … 0.6 Nm.
°
Tightening torque for M12 protective caps: 0.6 Nm … 0.7 Nm.
b
Mount the optics cover.
NOTICE
Plug in the system plug carefully.
Do not force it.
The contacts may break off or bend if too much force is used.
Replacing the system plug:
Tool required:
•
TX20 Torx wrench
Approach
1.
Make sure that the environment is clean and clear of fog, moisture, and dust.
2.
Disconnect the connecting cables the system plug.
3.
If necessary: move the safety laser scanner to a clean location.
4.
Unscrew screws in the defective system plug and remove the system plug from the
safety laser scanner.
MOUNTING INSTRUCTIONS
16
M O U N T I N G I N S T R U C T I O N S | microScan3 – PROFINET (RJ45, SCRJ)
8021229/13IA/2019-05-16 | SICK
Subject to change without notice