4
CALIBRATION PROCEDURE
35
CALIB
R
ATION PR
OCE
DURE
Chapter
4
4-8.
Digital Zero Limit
This is a load limit to allow digital zero.
When digital zero is executed, if the difference from the zero calibration point is larger than the
setting value, an error will result, and the indicated value will not be zeroed.
Setting range: 0 - 99999
How to set
1.
Press the [SET.] button. (refer to page 32)
2.
Press the [Calibration] button. (refer to page 32)
3.
Select page 2,
and press the [DZ
Limit] button.
4.
Enter the digital
zero limit value by the numerical keys
and determine with the [OK] bu
tton.
4-9.
Increment
(This step may be omitted if there is no change.)
Set the increment of the indicated value.
1 , 2 , 5 , 10 , 20 , 50 , 100 (The display changes by the decimal place.)
How to set
1.
Press the [SET.] button. (refer to page 32)
2.
Press the [Calibration] button. (refer to page 32)
3.
Select page 2,
and press the
[Increment]
button.
4.
Select the
increment and determine with the
[OK] button.
SET.
Calibration
Page 2
→
→
SET.
Calibration
Page 2
→
→