CALIBRATION PROCEDURE
32
CALIBRATION PROCEDURE
Chapter
4
4-3.
Excitation Voltage
Select the bridge voltage supplied to the strain gauge sensor.
The bridge voltage can be selected from 2.5V and 10V.
After this setting, be sure to perform calibration.
How to set
4-4.
Zero Calibration
Set the zero point in no-load condition.
How to set
1.
Press the [SET.] button.
2.
Press the [Calibration] button.
3.
Press the [Zero Cal.] button.
4.
Press [OK] button after confirming no-
load was applied to the sensor.
SET.
Calibration
Page 1
→
→
SET.
Calibration
Page 1
→
→