4
CALIBRATION PROCEDURE
34
CALIBRATION PROCEDURE
Chapter
4
■
Registration method at decimal place
Example)
The indicated value is made “200.0” by Equivalent Input Calibration.
It is a specification to which the decimal point is input with the numerical keys.
Please input it in the image that operates the calculator.
Please put it again from the start pushing
when you correct it.
4-7.
Digital Offset
By using the digital offset function, the value obtained by subtracting the set value from the
indicated value is displayed. This function is convenient when zero cannot be obtained with no load
for some reason or for offsetting. When not using, set "0".
(Displayed value)=(Actual indicated value)-(Offset value)
How to set
1.
Press the [SET.] button. (refer to page 32)
2.
Press the [Calibration] button. (refer to page 32)
3.
Select page 2,
and press the
[Digital Offset]
button.
4.
Enter the digital
offset value by the numerical keys and
determine with the [OK] button.
→
→
The display of the decimal point
→
→
Press
of the input value moves.
・
The decimal point ten keys are registered pushing at the end after the
numerical value is input when the decimal point is lost.
・
All set values concerning the load synchronize with the decimal place of the
indicated value (capacity).
Point
Setting range
:
-
99999 - 99999
SET.
Calibration
Page 2
→
→