This indicates error or operation status.
If there are multiple messages to be displayed, the priority message is displayed.
( 1 ) ROM : Is displayed when a ROM error has occurred.
( 2 ) RAM : Is displayed when a RAM error has occurred.
( 3 ) PARAMETER : Is displayed when a parameter memory error has occurred.
( 4 ) HOME POSITION : Is displayed when the optical axis is not adjusted correctly.
( 5 ) LEAKAGE : Is displayed when eluent leakage has occurred.
( 6 ) TEMP. ERROR : Is displayed when the temperature sensor has a problem.
( 7 ) OVER HEAT : Is displayed when the temperature is excessively high.
( 8 ) OPT. BALANCE : Is displayed when Auto Zero does not function correctly.
( 9 ) INTENSITY : Is displayed when the light intensity is low due to bubbles or dirt.
(10) TEMP. UNSTABLE : After the detector is started, this remains displayed
until the temperature remains within
1
C of the set
temperature for 10 minutes.
(11)Cannot set the LAN unit / Push the Enter Key : Is displayed when a LAN setting error
has occurred.
<Parameter setting screen>
To show the currently set parameters in order, press the key or the key
while the parameter setting screen is displayed.
< Validation screen >
To show the Validation screen in order, press the key or the key
while the practice screen of each validation and the result is displayed.
To change parameters, perform the following procedures:
(1) Display the relevant parameter.
(2) Press the key.
The underlined digits will flash and the parameter can be changed.
(3) Press the key or the key to change the parameter with
reference to Table 5-1.
(4) Press key to finalize the parameter.
(5) Press key to return to the monitor screen.
□