10.  Performance Test 

The column performance can be checked by calculating the number of theoretical plates under the 
following conditions: 
 
1) Specimen  

 

 

3% Acetic acid 

2) Injection amount  

 

5 µL 

3) Mobile phase  

 

0.1% Phosphoric acid aqueous solution 

4) Flow rate  

 

 

1 mL/min 

5) Column temperature  

40°C 

6) Detector  

 

 

High-sensitivity refractive index detector such as Shodex RI 

7) Calculation formula   

NTP = 5.54 x (Rt/W)

2

 

where NTP: Number of theoretical plate, Rt: Retention time, W: Peak half width 

 

 

 
 

11.  Attention 

1) Do not remove the end fittings of the column under any circumstances. 
2) Do not make a strong impact on the column: such as hitting or dropping on the floor. 
3) Replace the solvent in the LC system with the eluent to be used before connecting the column. 
4) Connect the column so that the flow direction corresponds to the arrow mark on the tag. 
5) When the column is not used for a month or more, replace the in-column solvent with the shipping 
solvent, remove it from the LC system, close each end with a stopper, and store it at controlled room 
temperature. 
6) Filtrate the sample with a disposable filter (0 .45 µm) to prevent deterioration by adsorbing insoluble 
matters. 
7)  Contact  Shodex  website  (www.shodex.de)  or  Shodex  partners  regarding  product  and  analysis 
applications. 
 

Summary of Contents for RSpak KC-811

Page 1: ... Shodex RSpak KC 811 and KC 811 6E Europe Middle East Africa Russia www shodex de Sales Office Showa Denko Europe GmbH Shodex Business Konrad Zuse Platz 3 81829 Munich Germany Technical Support info shodex de 49 0 89 93 99 62 37 Quotes and Orders order shodex de 49 0 89 93 99 62 34 ...

Page 2: ...oblems related to losing your health or leaking Attention Use the column within the regular range of flow rate pressure and temperature There is a danger of deteriorating the performance when it is handled beyond the permissible range even for a short time See the Section 4 Usable Conditions for the permissible ranges 3 Specifications Product code Product name Functional group Size mm Particle siz...

Page 3: ... amount up to 20 to the mobile phase to reduce the absorptivity 6 Mounting and Start up 1 Before mounting the column on the liquid chromatograph replace the solvent in the chromatograph with the mobile phase Caution In replacing a water insoluble organic solvent in the chromatograph replace it first with a water soluble organic solvent such as acetone or ethanol and then replace the water soluble ...

Page 4: ...ounting and Storage 1 Set the flow rate at 0 2 mL min turn off the heater and let the column stand as is until it cools down to room temperature 2 Stop the pump and dismount the column 3 Cap both ends of the column put it back in the same case in which it was delivered from the manufacturer and store the case in a place where the temperature does not fluctuate more than 5 C Caution i Do not remove...

Page 5: ...not remove the end fittings of the column under any circumstances 2 Do not make a strong impact on the column such as hitting or dropping on the floor 3 Replace the solvent in the LC system with the eluent to be used before connecting the column 4 Connect the column so that the flow direction corresponds to the arrow mark on the tag 5 When the column is not used for a month or more replace the in ...

Page 6: ...of the following cases 3 1 If the Shodex Column is not used in accordance with the operating manual 3 2 If the Shodex Column is remodelled by anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex Column is disposed of 3 4 If the Shodex Column is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex Column is not co...

Reviews: