Shodex OHpak SB-2000 Series Operation Manual Download Page 3

ver.211025E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

· Shear degradation is more likely to occur in larger molecular weight compounds. When 

shear degradation occurs, the measured molecular weight may be lower than the actual 
value or it may influences the repeatability of the analysis. If shear degradation is 
suspected, use a lower flow rate.   

 

5. Eluent Preparation 

(1) Degas the eluent fully to prevent the formation of air bubbles.   

 

(2) Presence of small debris or insoluble substances may result in deterioration of the column and/or they appear 

as noise on the chromatograms. Filter the eluent with a 0.45-

μ

m disposable filter to prevent the problems. 

 

 

Attention!

   

· Whenever water is required, use ultra-pure water freshly generated by a water purification 

system or water from a newly opened HPLC grade distilled water bottle. Use HPLC grade 
organic solvent whenever possible. Solvents left in an opened bottle for a long time should 
not be used. The content may have been changed, absorbed moisture, or has been 
contaminated. 

 

· Always use freshly prepared solvents. Solvents stored for a long time may have changed 

their compositions and may influence elution patterns and/or damage the column.

   

 

 

       

Note  

 

 

· Use of on-line degasser is recommended. 

 

6. Sample Preparation 

(1) If possible, use the eluent for analysis to dissolve or dilute samples. If this is difficult, use a solvent which 

has a composition that is as close as possible to the eluent's composition, but which fully dissolves or dilutes 
the sample. 

 

(2) Filter the sample solution using disposable 0.45-

μ

m filter to prevent the column from clogging or deteriorating. 

 

(3) To prepare a sample with molecular weight larger than 1,000,000, first allow the sample to stand in the eluent 

of analysis for 1 day until it becomes fully swollen. Next, slowly agitate the sample solution to completely 
dissolve the sample. Be careful as aggressive agitation can cut the polymer chains of the analyte.   

 

(4) Recommended injection volume is 

250 - 1,000 

μ

L per column.   

 

(5) Viscosity of high molecular weight compound is largely influenced by its molecular weight and concentration. 

Samples with high viscosity cause peak broadening and elution delay, and this makes it difficult to obtain 
their accurate molecular weight distributions. In general, the larger the molecular weight of the compound, 
the higher its viscosity becomes. To suppress the influence from high viscosity, it is recommended to lower 
the sample concentration. Please use the below table as a reference when preparing samples for molecular 
weight distribution analyses.   

 

Molecular Weight Range 

Optimal Concentration (w/v) 

 5,000 

 1.0 % 

5,000 -   25,000 

 0.5 % 

25,000 -  200,000 

  

 0.25 % 

200,000 - 2,000,000 

 0.1 % 

2,000,000 

 

 

 0.05 % 

 

       

Note   

 

· Use of guard column is recommended to protect the analytical column. 

 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for OHpak SB-2000 Series

Page 1: ...ries is a size exclusion chromatography column used with aqueous solvents The column series is suitable for the preparations of various water soluble high molecular weight polymers proteins and or oligomers and molecular weight distribution analysis of high molecular weight polymers The columns with different exclusion limits are available Please select columns with target molecular weight ranges ...

Page 2: ...d to the eluent However since surfactants tend to remain on the column solvent replacement after their use takes a longer time than replacement from general solvents The replacement time can be shorten by using 30 to 50 v v methanol Organic solvents cannot be used for SB 2002 Attention Use the column within above stated flow rate pressure and temperature ranges Using the column outside the given r...

Page 3: ...t for analysis to dissolve or dilute samples If this is difficult use a solvent which has a composition that is as close as possible to the eluent s composition but which fully dissolves or dilutes the sample 2 Filter the sample solution using disposable 0 45 μm filter to prevent the column from clogging or deteriorating 3 To prepare a sample with molecular weight larger than 1 000 000 first allow...

Page 4: ... Set the initial flow rate at less than 1 0 mL min and start the system If using the column at an elevated temperature keep a low flow rate until the temperature of the column reaches to the set temperature and then gradually increase the flow rate to the desired 4 Multiple number of columns can be connected in series When connecting multiple columns with each having different exclusion limits set...

Page 5: ...mn first remove the guard column and check the performance of the analytical column alone If the problem is solved most likely the cause is from the guard column In this case clean the guard column If the problem is not solved by removing the guard column clean both guard and analytical columns Make sure to clean the guard and the analytical columns separately When washing the column let the washi...

Page 6: ...late Number N 2 Asymmetry Factor Fas 10 Additional Warnings 1 Do not remove end fittings 2 Do not make a strong impact on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support of...

Page 7: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Reviews: