Operation Manual
Shodex
OHpak
SB-2000 series
(
Please read this manual carefully to achieve the best and consistent column performance for a long time
)
Important Handling Instructions
Caution!
· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column
and understand their proper handling methods to prevent potential health hazards or death
from occurring.
· Please wear appropriate personal protective equipment such as lab goggles and gloves
when handling organic solvents and acid and alkaline reagents. Avoid any direct physical
contact to prevent chemical injuries.
Before Using the Column
(1) Please visually inspect the package and outside of the column for any damage.
(2) Please check if product name and serial number written on the column package, column adhesive label, and
enclosed Certificate of Analysis (CoA) are matching and correct.
1. Introduction
Thank you for purchasing the Shodex™ product. Shodex
OHpak
SB-2000 series is a size exclusion
chromatography column used with aqueous solvents. The column series is suitable for the preparations of
various water-soluble high molecular weight polymers, proteins, and/or oligomers, and molecular weight
distribution analysis of high molecular weight polymers. The columns with different exclusion limits are available.
Please select columns with target molecular weight ranges (see below) that meet the molecular weight range
of your samples.
2. Column Components
3. Column Specifications
Product
Code
Product Name
Column Size (mm)
Particle
Size (
μ
m)
Theoretical
Plate Number
(Per Column)
Exclusion Limit
*1
I.D. Length
F6516011
OHpak SB-2002
20.0 300 15
≥
9,000
1,000
F6516012
OHpak SB-2002.5
20.0 300 10
≥
12,000
10,000
F6516013
OHpak SB-2003
20.0 300 10
≥
12,000
100,000
F6516014
OHpak SB-2004
20.0 300 18
≥
12,000
1,000,000
F6516015
OHpak SB-2005
20.0 300 20
≥
12,000
(4,000,000)
*2
F6516016
OHpak SB-2006
20.0 300 20
≥
12,000
(20,000,000)
*2
F6516017
OHpak SB-2006M
20.0 300 20
≥
12,000
(20,000,000)
*2
F6709555
OHpak SB-G 8B
8.0
50 18 (Guard
Column)
-
SB-806M HQ is a mixed-gel column, suitable for the analysis of wide molecular weight ranges.
*1
Reference value only / Measured with pullulan
*2
Estimated value