Shodex IEC SP-FT 4A Operation Manual Download Page 2

ver.211110E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

4.

 

Usable Conditions 

 

* Total pressure of column and HPLC system 

 

Usable solvents are listed below.   
(1)

 

The standard eluents are buffers and aqueous solutions of salt such as sodium chloride, potassium chloride, 
sodium sulfate, and potassium sulfate. Salts may also be added to the buffers. The recommended salt 
concentration ranges are 0.1 - 1 M. The total salt concentration can be up to 1.5 M.   

Select a buffer which has negative charge ion and has a large buffering capacity at the set pH. 

 

pH Range 

Buffer 

3.8 - 4.3 

Sodium formate 

4.3 - 4.8 

Sodium succinate 

4.8 - 5.2 

Sodium acetate 

5.0 - 6.0 

Sodium malonate 

5.5 - 7.0 

2-(N-morpholino)ethanesulfonic acid (MES) 

6.7 - 7.6 

Sodium phosphate 

6.8 - 8.2 

2-[4-(2-Hydroxyethyl)-1-piperazinyl]-ethanesulfonic acid (HEPES) 

7.7 - 9.1 

N,N-Bis(2-hydroxyethyl) glycine (BICINE) 

8.6 - 10.0 

2-(cyclohexylamino)ethanesulphonic acid (CHES) 

 

(2) Up to 20 % (v/v) acetonitrile can be added.   

 

 

Attention!

   

· Use the column within above stated flow rate, pressure, and temperature ranges. Using 

the column outside the given range may damage the column and lower its performance.   

 

· When using a mixture of buffer (or aqueous solution of salt) and organic solvent, make 

sure there is no precipitation of salt. 

 

· When using highly corrosive salts such as sodium chloride, wash out the salts at the end 

of analysis. The metal parts of the devices may rust.

   

 

 

·

 

Column pressure is influenced by the eluent composition, flow rate, and column  

temperature. When changing the eluent compositions, adjust the flow rate and column 
temperature so that the column pressure remains below the usable maximum pressure.   

 

5. Eluent Preparation 

(1) Degas the eluent fully to prevent the formation of air bubbles.   

 

(2) Presence of small debris or insoluble substances may result in deterioration of the column and/or they appear 

as noise on the chromatograms. Filter the eluent with a 0.45-

μ

m disposable filter to prevent the problems. 

 

 

Attention!

   

· Whenever water is required, use ultra-pure water freshly generated by a water purification 

system or water from a newly opened HPLC grade distilled water bottle. Use HPLC grade 
organic solvent whenever possible. Solvents left in an opened bottle for a long time should 
not be used. The content may have been changed, absorbed moisture, or has been 
contaminated. 

 

· Always use freshly prepared solvents. Solvents stored for a long time may have changed 

their compositions and may influence elution patterns and/or damage the column.

   

 

 

       

Note  

 

 

· Use of on-line degasser is recommended. 

 
 
 
 

Product Name 

Flow Rate (mL/min) 

Maximum Pressure (MPa) 

(System Pressure*)

 

pH 

Range 

Temperature 

Range (°C) 

Recommended Maximum 

IEC SP-FT 4A 

1.0 - 2.0 

3.0 

20 

2 - 12 

5 - 45 

Summary of Contents for IEC SP-FT 4A

Page 1: ...and serial number written on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex IEC SP FT 4A is a strong cation exchange column filled with non porous hydrophilic polymer based gel modified with sulfopropyl group Its small hydrophobic adsorption makes the column suitable for analyzin...

Page 2: ...lt When using highly corrosive salts such as sodium chloride wash out the salts at the end of analysis The metal parts of the devices may rust Column pressure is influenced by the eluent composition flow rate and column temperature When changing the eluent compositions adjust the flow rate and column temperature so that the column pressure remains below the usable maximum pressure 5 Eluent Prepara...

Page 3: ... 2 Column Installation 1 Connect the column to LC system by following the flow direction arrow indicated on the column name tag If guard column is used position the guard column in front before the inlet of the analytical column 2 Make sure to insert the tubing all the way to the end fitting and secure it with the male nut It is important that there is no extra space between the tubing and the col...

Page 4: ...usable maximum pressure 7 4 Column Cleaning Problems in peak shapes and elution time changes or elevated column pressure etc are often caused by insoluble or adsorbing components present in the eluent and reagents being deposited inside the column These problems may be resolved by cleaning the column Make sure to clean the guard and the analytical columns separately When washing the column let the...

Page 5: ...late Number N 2 Asymmetry Factor Fas 10 Additional Warnings 1 Do not remove end fittings 2 Do not make a strong impact on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support of...

Page 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Reviews: