6. Sample Preparation 

(1) If possible, use the eluent for analysis to dissolve or dilute samples. If this is difficult, use a solvent which 

has a composition that is as close as possible to the eluent's composition, but which fully dissolves or dilutes 
the sample. When gradient elution is used, it is recommended to use the initial eluent to prepare the sample.     

 

(2) Filter the sample solution using disposable 0.45-

μ

m filter to prevent the column from clogging or deteriorating. 

 

(3) Recommended sample injection volume is less than 100 

μ

L per column. 

 

(4) When a sample contains lipid, make sure to remove them prior to the injection.   
 

7. Column Usage Procedure 

 

7.1 HPLC System Preparation 

Wash entire LC system prior to the column installation, including all flow-lines and sample loop by switching the 
valve, and then replace the washing solution with the eluent to be used. If desired new eluent has low 
miscibility/solubility to the eluent of previous analysis, first use the eluent that is miscible/soluble to both eluents, 
and then replace it with the desired eluent.   

 

e.g. When replacing chloroform to buffer, first run methanol and then introduce water, followed by the desired 

buffer. 

 

 Attention! 

 

· If the eluent left in the system is not compatible with the column to be used, it may damage 

the column. 

 

 

     · A drastic change in the eluent compositions may remove substances adsorbed on the 

system and they may enter and deteriorate the column. 

 

7.2 Column Installation

 

 

(1) Connect the column to LC system by following the “flow direction arrow” (

) indicated on the column name 

tag.  

 

(2) Make sure to insert the tubing all the way to the end fitting and secure it with the male nut. It is important that 

there is no extra space between the tubing and the column side of the end fitting. Presence of an extra space 
will let the sample to spread out and may result in wide peaks. 

 

 

 

(3)

 

Set the initial flow rate at less than 0.5 mL/min and start the system. If using the column at an elevated 

temperature, keep a low flow rate until the temperature of the column reaches to the set temperature, and 
then gradually increase the flow rate to the desired.   

 

 Caution! 

 

 

· Verify that there is no solvent leak. It may cause electronic leakage, rust, and/or chemical 

injury. 

 

 

Attention!

   

· Make sure not to let air bubbles enter the column while installing the column. The air 

bubbles may damage the column.   

 

· When restarting the system after column installation or after holding the eluent flow, start 

the system at less than 0.5 mL/min. A rapid increase in pressure can damage the column.   

 

· If the column was used at an elevated temperature, lower the flow rate to less than      

0.5 mL/min at the end of analysis. Then, turn off the column oven, and let the column 
temperature return to room temperature before stopping the pump. If the pump was 
stopped while the eluent inside the column was still hot, as the eluent temperature 
decreases, its volume also decreases. This may result in creating an empty space in the 
column and deteriorates the column. 

 

       

Note  

 

 

· It is recommended to set the pump limiter to avoid exceeding the maximum pressure.   

 

Summary of Contents for IEC SP-825

Page 1: ...rial number written on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex IEC SP 825 is a strong cation exchange column filled with hydrophilic polymer based gel modified with sulfopropyl group It is an ideal column for analyzing various proteins peptides and components of nucleic ac...

Page 2: ...orrosive salts such as sodium chloride wash out the salts at the end of analysis The metal parts of the devices and or the columns may rust Column pressure is influenced by the eluent composition flow rate and column temperature When changing the eluent compositions adjust the flow rate and column temperature so that the column pressure remains below the usable maximum pressure 5 Eluent Preparatio...

Page 3: ...orate the column 7 2 Column Installation 1 Connect the column to LC system by following the flow direction arrow indicated on the column name tag 2 Make sure to insert the tubing all the way to the end fitting and secure it with the male nut It is important that there is no extra space between the tubing and the column side of the end fitting Presence of an extra space will let the sample to sprea...

Page 4: ...c are often caused by insoluble or adsorbing components present in the eluent and reagents being deposited inside the column These problems may be resolved by cleaning the column When washing the column let the washing solution flow from the column outlet go directly into the waste container and not let the solution go through the detector Cleaning Method 1 Insoluble components that block the colu...

Page 5: ...late Number N 2 Asymmetry Factor Fas 10 Additional Warnings 1 Do not remove end fittings 2 Do not make a strong impact on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support of...

Page 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Reviews: