Shodex IC SI-35 4D Operation Manual Download Page 2

ver.211104E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

4.

 

Usable Conditions 

 

Usable solvents are listed below. 
(1) In general, sodium carbonate solution

 

or a mixture of sodium carbonate and sodium bicarbonate solutions 

is used. 

 

(2) The standard eluent is 3.6-mM sodium carbonate solution. The higher the concentration of sodium carbonate 

solution, the faster the elution of ions will be.   

 

(3) Up to 10 % (v/v) acetonitrile or methanol can be added.   

 

 

Attention!

   

· Use the column within above stated flow rate, pressure, and temperature ranges. Using 

the column outside the given range may damage the column and lower its performance.     

 

·

 

Column pressure is influenced by the eluent composition, flow rate, and column  

temperature. When changing the eluent compositions, adjust the flow rate and column 
temperature so that the column pressure remains below the usable maximum pressure.   

 
5. Eluent Preparation 

(1) Degas the eluent fully to prevent the formation of air bubbles.   

 

(2) Presence of small debris or insoluble substances may result in deterioration of the column and/or they appear 

as noise on the chromatograms. Filter the eluent with a 0.45-

μ

m disposable filter to prevent the problems. 

 

 

Attention!

   

· Whenever water is required, use ultra-pure water freshly generated by a water purification 

system or water from a newly opened HPLC grade distilled water bottle. Use HPLC grade 
organic solvent whenever possible. Solvents left in an opened bottle for a long time should 
not be used. The content may have been changed, absorbed moisture, or has been 
contaminated. 

 

· Always use freshly prepared solvents. Solvents stored for a long time may have changed 

their compositions and may influence elution patterns and/or damage the column.

     

 

 

       

Note  

 

 

· Use of on-line degasser is recommended. 

 
6. Sample Preparation

   

 

 

(1) Filter the sample solution using disposable 0.45-

μ

m filter to prevent the column from clogging or deteriorating. 

 

(2) Recommended sample injection volume is less than 50 

μ

L per column.   

 

(3)

 

For the samples containing proteins, remove the proteins prior to the sample injection. The use of an 

ultrafiltration membrane is recommended. 

 

(4) For the samples containing hydrophobic impurities, remove the impurities prior to the sample injection. The 

use of a solid phase extraction is recommended. 

 

       

Note   

 

· Use of guard column is recommended to protect the analytical column. 

 
 
 
 
 
 
 
 
 
 
 
 
 

Product Name 

Flow Rate (mL/min)

 

Maximum 

Pressure (MPa) 

(Per Column)

 

pH 

Range

 

Temperature (°C)

 

Recommended

 

Maximum

 

Recommended

  Range 

IC SI-35 4D 

0.6

 

0.8

 

15

 

3 - 12

 

45

 

20 - 60

 

IC SI-95G 

- -  - 

Summary of Contents for IC SI-35 4D

Page 1: ...me and serial number written on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex IC SI 35 4D is a suppressor type ion chromatography column for analysis of anions using sodium carbonate eluents The column is suitable for the rapid analysis of oxyhalides including chlorite chlorate ...

Page 2: ...t the problems Attention Whenever water is required use ultra pure water freshly generated by a water purification system or water from a newly opened HPLC grade distilled water bottle Use HPLC grade organic solvent whenever possible Solvents left in an opened bottle for a long time should not be used The content may have been changed absorbed moisture or has been contaminated Always use freshly p...

Page 3: ...t and may result in wide peaks 3 Set the initial flow rate at less than 0 3 mL min and start the system If using the column at an elevated temperature keep a low flow rate until the temperature of the column reaches to the set temperature and then gradually increase the flow rate to the desired Caution Verify that there is no solvent leak It may cause electronic leakage rust and or chemical injury...

Page 4: ...om the column outlet with flow rate at less than half of the recommended flow rate 2 Follow below cleaning steps for adsorbing components For an efficient cleaning reverse the flow direction and use the flow rate at 0 3 mL min Method 1 Adsorption of hydrophilic compounds Washing Solution Cleaning Time 1 H2O 15 min 2 10 x Concentrated eluent 60 min 3 H2O 15 min 4 Eluent Return the column flow to no...

Page 5: ...late Number N 2 Asymmetry Factor Fas 10 Additional Warnings 1 Do not remove end fittings 2 Do not make a strong impact on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support of...

Page 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Reviews: