Shodex C18U 2B Operation Manual Download Page 4

ver.210601E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

· If the column was used at an elevated temperature, lower the flow rate to less than half of 

the recommended flow rate at the end of analysis. Then, turn off the column oven, and let 
the column temperature return to room temperature before stopping the pump. If the pump 
was stopped while the eluent inside the column was still hot, as the eluent temperature 
decreases, its volume also decreases. This may result in creating an empty space in the 
column and deteriorates the column. 

 

       

Note  

 

 

· It is recommended to set the pump limiter to avoid exceeding the maximum pressure.   

 

7.4 Solvent Exchange 

(1) When replacing the solvent, start the system at less than 0.2 mL/min. Recommended solvent volume to 

introduce at each step is 3 to 5 times of the column volume. 

 

(2) Check miscibility/solubility of the desired new solvent and the solvent currently filled in the column. 

 

(3)

 

When replacing the current solvent with a solvent with low miscibility/solubility to the current solvent, first 
use a solvent that is miscible/soluble to both eluents, and then replace it with the new solvent. 

 

e.g. When replacing highly concentrated buffer solution or salt solution to water/acetonitrile, first run water 

and then replace it with the eluent. 

 

(4) When using a gradient method, changes in the eluent compositions may increase the column backpressure. 

Adjust the flow rate and column temperature so that the column backpressure remains below the usable 
maximum pressure.

 

 

7.5 Column Cleaning 

Problems in peak shapes and elution time changes or elevated column pressure etc. are often caused by 
insoluble or adsorbing components present in the eluent and reagents being deposited inside the column. These 
problems may be resolved by cleaning the column.   
Make sure to clean the guard and the analytical columns separately. When washing the column, let the washing 
solution flow from the column outlet go directly into the waste container and not let the solution go through the 
detector. 

 

<Cleaning Method> 
(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., 

introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate. 

 

(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction 

and use the flow rate less than 0.2 mL/min. Introduce 5 to 10 times column volume of the washing solvent.   

 

Method 1: After using eluent not containing buffer salts/additives 
Wash the column with solution containing a higher ratio of organic solvent that constitutes the eluent. A 
cleaning solution containing THF might be effective when removing highly hydrophobic (lipid-soluble) 
substances that are adsorbed onto the gel. 

 

Method 2: After using mobile phase containing buffer salts/additives 

 

First replace with a water/organic solution containing no buffer salts or additives (A ratio of water to organic 
solvent should be set at the same proportion as eluent). Then flush the column in accordance with the method 
described above. 

 

 

Attention!

   

· Flushing with 100% water after using the column around the pH limit might shorten column 

life, so flush the column with water/organic solution.   

 

8. Column Storage

 

Remove the column from the system after replacing the in-column solvent with the shipping solvent. Securely 
tighten the end caps and store the column at a location with stable temperature (a cool and dark space is 
recommended). Refer to section 7.4 Solvent Exchange for how to replace the eluent. 

 

 

Attention!

   

· Never allow inside of the column to dry. It can damage the column.     

 
 
 
 
 
 

Summary of Contents for C18U 2B

Page 1: ...tten on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex C18U series is reverse phase chromatography column in which octadecyl group is bonded to organic and inorganic hybrid gel Compared to conventional silica based reversed phase columns it has excellent separation performance an...

Page 2: ...n 1 Degas the eluent fully to prevent the formation of air bubbles 2 Presence of small debris or insoluble substances may result in deterioration of the column and or they appear as noise on the chromatograms Filter the eluent with a 0 2 μm disposable filter to prevent the problems Attention Whenever water is required use ultra pure water freshly generated by a water purification system or water f...

Page 3: ...is miscible soluble to both eluents and then replace it with the desired eluent e g When replacing highly concentrated buffer solution or salt solution to water acetonitrile first run water and then replace it with the eluent Attention If the eluent left in the system is not compatible with the column to be used it may damage the column A drastic change in the eluent compositions may remove substa...

Page 4: ...de the column These problems may be resolved by cleaning the column Make sure to clean the guard and the analytical columns separately When washing the column let the washing solution flow from the column outlet go directly into the waste container and not let the solution go through the detector Cleaning Method 1 Insoluble components that block the column inlet may be removed by reversing the flo...

Page 5: ...t on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support office https www shodex com en support_office list ...

Page 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Reviews: