Shodex Asahipak NH2P-40 2E Operation Manual Download Page 4

ver.211215E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

7.3 Solvent Exchange 

(1) When replacing the solvent, start the system at less than 0.1 mL/min. Recommended solvent volume to 

introduce at each step is 3 to 5 times of the column volume. 

 

(2) Check miscibility/solubility of the desired new solvent and the solvent currently filled in the column. 

 

(3)

 

When replacing the current solvent with a solvent with low miscibility/solubility to the current solvent, first 
use a solvent that is miscible/soluble to both eluents, and then replace it with the new solvent. 

 

e.g. When replacing water/acetonitrile to a highly concentrated buffer solution, first run water and then 

replace it with the buffer solution.   

 

(4) When using a gradient method, changes in the eluent compositions may increase the column backpressure. 

Adjust the flow rate and column temperature so that the column backpressure remains below the usable 
maximum pressure. 

 

 

Attention!

   

· When replacing the eluent from water/acetonitrile to a mixture of buffer solution and organic 

solvent, take a pre-equilibrium step: First run 10 - 20 column volumes of the buffer without 
organic solvent, and then replace it with the desired eluent. If replacing to a buffer solution 
alone or a mixture of water and ethanol, pre-equilibrium step is generally not required. 

 

·

 

When replacing the eluent from a mixture of buffer solution and organic solvent to 

water/organic solvent, first wash the column with alkaline solution. Please refer to “7.4 
Column Cleaning Method 1”. 

 

7.4 Column Cleaning 

Problems in peak shapes and elution time changes or elevated column pressure etc. are often caused by 
insoluble or adsorbing components present in the eluent and reagents being deposited inside the column. These 
problems may be resolved by cleaning the column. When washing the column, let the washing solution flow 
from the column outlet go directly into the waste container and not let the solution go through the detector. 

 

<Cleaning Method> 
(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., 

introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate. 

 

(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction. 

Use the flow rate at 0.1 mL/min. 

 

 

Method 1: General cleaning 

Washing Solution 

Cleaning Time 

1 H

2

O  

 10 min 

0.1 M NaOH aq. solution 

120 min 

3 H

2

 20 min 

4 Eluent 

 60 min 

 

Method 2: Adsorption of metal ions or in case Method 1 did not work   

Washing Solution 

Cleaning Time 

1 H

2

O  

 10 min 

0.1 M HNO

3

 aq. solution 

120 min 

3 H

2

 10 min 

0.1 M NaOH aq. solution 

120 min 

5 H

2

 20 min 

6 Eluent 

 60 min 

 
 

 

Attention!

   

· Keeping acidic or alkaline solution in the column for a long time will lead to the column 

deterioration. Complete the suggested column cleaning steps all the way up to the eluent 
replacement at once and do not hold the procedure in middle way.   

 

 

· Strong alkaline solvents such as 0.1-M aqueous sodium hydroxide solution can damage 

the detector cell. Do not connect the detector while regenerating the column and collect 
elute directly from the column outlet to the waste container. 

Summary of Contents for Asahipak NH2P-40 2E

Page 1: ... Please visually inspect the package and outside of the column for any damage 2 Please check if product name and serial number written on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex Asahipak NH2P series is polymer based chromatography columns modified with amino group The colu...

Page 2: ...as noise on the chromatograms Filter the eluent with a 0 45 μm disposable filter to prevent the problems Attention Whenever water is required use ultra pure water freshly generated by a water purification system or water from a newly opened HPLC grade distilled water bottle Use HPLC grade organic solvent whenever possible Solvents left in an opened bottle for a long time should not be used The con...

Page 3: ...t the tubing all the way to the end fitting and secure it with the male nut It is important that there is no extra space between the tubing and the column side of the end fitting Presence of an extra space will let the sample to spread out and may result in wide peaks 3 Set the initial flow rate at 0 1 mL min and start the system If using the column at an elevated temperature keep a low flow rate ...

Page 4: ...Cleaning Method 1 7 4 Column Cleaning Problems in peak shapes and elution time changes or elevated column pressure etc are often caused by insoluble or adsorbing components present in the eluent and reagents being deposited inside the column These problems may be resolved by cleaning the column When washing the column let the washing solution flow from the column outlet go directly into the waste ...

Page 5: ...olumn Inspection Inspection method is described in the Certificate of Analysis CoA Theoretical Plate Number N and Asymmetry Factor Fas were calculated using the below equations 1 Theoretical Plate Number N 2 Asymmetry Factor Fas 10 Additional Warnings 1 Do not remove end fittings 2 Do not make a strong impact on the column Do not drop or hit the column on a hard surface Refer to the Shodex website...

Page 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Reviews: