· Shear degradation is more likely to occur in larger molecular weight compounds. When 

shear degradation occurs, the measured molecular weight may be lower than the actual 
value or it may influences the repeatability of the analysis. If shear degradation is 
suspected, use a lower flow rate. 

 

5. Eluent Preparation 

(1) Degas the eluent fully to prevent the formation of air bubbles.   

 

(2) Presence of small debris or insoluble substances may result in deterioration of the column and/or they appear 

as noise on the chromatograms. Filter the eluent with a 0.45-

μ

m disposable filter to prevent the problems. 

 

 

Attention!

   

· Whenever water is required, use ultra-pure water freshly generated by a water purification 

system or water from a newly opened HPLC grade distilled water bottle. Use HPLC grade 
organic solvent whenever possible. Solvents left in an opened bottle for a long time should 
not be used. The content may have been changed, absorbed moisture, or has been 
contaminated. 

 

· Always use freshly prepared solvents. Solvents stored for a long time may have changed 

their compositions and may influence elution patterns and/or damage the column.

   

 

 

       

Note  

 

 

· Use of on-line degasser is recommended. 

 

6. Sample Preparation 

(1) If possible, use the eluent for analysis to dissolve or dilute samples. If this is difficult, use a solvent which 

has a composition that is as close as possible to the eluent's composition, but which fully dissolves or dilutes 
the sample. 

 

(2) Filter the sample solution using disposable 0.45-

μ

m filter to prevent the column from clogging or deteriorating. 

 

(3) Recommended injection volume is less than 1 mL per column.   

 

(4) Viscosity of high molecular weight compound is largely influenced by its molecular weight and concentration. 

Samples with high viscosity cause peak broadening and elution delay, and this makes it difficult to obtain 
their accurate molecular weight distributions. In general, the larger the molecular weight of the compound, 
the higher its viscosity becomes. To suppress the influence from high viscosity, it is recommended to lower 
the sample concentration. Please use the below table as a reference when preparing samples for molecular 
weight distribution analyses.   

 

Molecular Weight Range 

Optimal Concentration (w/v) 

 5,000 

 1.0 % 

5,000 -   25,000 

 0.5 % 

25,000 -  200,000 

  

 0.25 % 

200,000 - 2,000,000 

 0.1 % 

 

       

Note   

 

· Use of guard column is recommended to protect the analytical column. 

 

7. Column Usage Procedure 

 

7.1 HPLC System Preparation 

Wash entire LC system prior to the column installation, including all flow-lines and sample loop by switching the 
valve, and then replace the washing solution with the eluent to be used. If desired new eluent has low 
miscibility/solubility to the eluent of previous analysis, first use the eluent that is miscible/soluble to both eluents, 
and then replace it with the desired eluent.   

 

e.g. When replacing chloroform to water, first run methanol and then replace it with water. 

 

e.g. When replacing highly concentrated buffer solution or salt solution to water/acetonitrile, first run water and 

then replace it with water/acetonitrile. 

 

 Attention! 

 

· If the eluent left in the system is not compatible with the column to be used, it may damage 

the column. 

 

 

     · A drastic change in the eluent compositions may remove substances adsorbed on the 

system and they may enter and deteriorate the column. 

 

Summary of Contents for Asahipak GS-310 20G

Page 1: ...check if product name and serial number written on the column package column adhesive label and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex Asahipak GS 310 20G and GS 510 20G are polymer based size exclusion chromatography columns They work well with both aqueous and organic solvents 2 Column Components 3 Column Sp...

Page 2: ...ent It is recommended to dedicate the column for this specific use 4 Surfactants such as SDS and Brij 35 can also be added to the eluent However since surfactants tend to remain on the column solvent replacement after their use takes a longer time than replacement from general solvents The replacement time can be shorten by using 30 50 v v methanol Attention Use the column within above stated flow...

Page 3: ... clogging or deteriorating 3 Recommended injection volume is less than 1 mL per column 4 Viscosity of high molecular weight compound is largely influenced by its molecular weight and concentration Samples with high viscosity cause peak broadening and elution delay and this makes it difficult to obtain their accurate molecular weight distributions In general the larger the molecular weight of the c...

Page 4: ... at less than 3 0 mL min A rapid increase in pressure can damage the column If the column was used at an elevated temperature lower the flow rate to less than 3 0 mL min at the end of analysis Then turn off the column oven and let the column temperature return to room temperature before stopping the pump If the pump was stopped while the eluent inside the column was still hot as the eluent tempera...

Page 5: ... column outlet with flow rate at less than half of the recommended flow rate 2 Follow below cleaning steps for adsorbing components For an efficient cleaning reverse the flow direction and use the flow rate at less than 3 0 mL min Introduce 5 to 10 times column volume of the washing solvent Method 1 Adsorption of hydrophilic compounds when using aqueous eluent Introduce the eluent with a high pola...

Page 6: ...ct on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support office https www shodex com en support_office list ...

Page 7: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Reviews: