Measurement Display 

 
While  pressing  MEASURE  key,  aim  the  laser  beam  at  an  object  being
tested. MEAS is shown on the display to indicate that the instrument is in
measuring  mode.  Release  MEASURE  key  to  stop  measurement.  The
display reading is automatically frozen, with HOLD shown on the display. 
 
To display maximum or minimum reading; 
In the HOLD mode, press  MODE key to display the maximum reading.
MAX  is  shown  on  the  display.  Press  MODE  key  again  to  display  the
minimum reading. MIN is shown on the display. 
 
To display only the maximum or minimum reading during measurement; 
With MAX or MIN shown on the display, press MEASURE key. Updated
maximum or minimum reading is shown on the display. 
If  no  key  is  pressed  for  30  seconds  after  the  reading  is  frozen,  the
instrument automatically turns itself off. Pressing MEASURE key to turn
on the instrument cancels stored maximum or minimum reading. When the
batteries wear out, stored maximum or minimum reading is canceled as
well. 

 

 

Battery 

(1) Installing Batteries 

Press the battery compartment cover on its top and bottom as shown 

below to remove it. Install batteries in correct polarity as shown inside 

the battery compartment. 

 

(2) Replacing Batteries 

Replace  the  batteries  when  the 

low battery symbol appears on the 

display or the display blinks. 

 

  Note        Always replace both batteries. 

 

 

 

 

Settings 

(1) Emissivity 

Emissivity of an object being tested can be selected from DARK (0.95), 

BRIGHT (0.8) or FREE (variable from 0.3 to 1.90).   

When  the  instrument  is  turned  on,  emissivity  is  set  to  DARK.  A  bar 

appears under DARK, BRIGHT or FREE on the display to indicate which 

one  is  selected.  Press  ▲  key  to  change  selection.  When  FREE  is 

selected, emissivity is set to 1.00. To change this value; 

①  Press MEASURE key to turn on the instrument. 

②  While pressing ENT key, push MODE key. ALARM is shown on the 

display. 

③  Press  MODE key. 1.00 or the value set previously is shown on the       

display, with the least significant digit blinking. 

Press  ▲   or  ▼   key  to  change  the  blinking 

number and press ENT key. Repeat the same step 

on the other digits. 

④  Press MODE key to switch the instrument to measuring mode. 

 

(2) Alarm 

If readings exceed a certain value, the beeper sounds and ALARM blinks 

on the display. To set the alarm value; 

①  Press MEASURE key to turn on the instrument. 

②  While pressing ENT key, push MODE key. 

699  or  the  value  set  previously  is  shown  on  the 

display, with the least significant digit blinking. 

③  Press  ▲  and  ▼  key  to  change  the  blinking 

number and press ENT key. Repeat the same step on 

the other digits. 

To set 2-digit value, display 

0

 on the most significant digit. 

To set a negative value, display 

on the most significant digit. 

④  Press MODE key to switch the instrument to measuring mode. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

     

 

 

 

 

 

 

 

 

     

 

 

 

 

 

 

Emissivity 
DARK    :0.95 
BRIGHT:0.80 
FREE    :0.3 to 1.90 variable 
 
 
 
 
 
Displays temperature 
readings, emissivity 
value or alarm value 
 
 
 
 
 
 
 
 
 
 
MEAS:Displayed during measurement 
HOLD:Displayed during hold mode 

Low Battery Symbol 
Blinks  when  batteries  wear
out 
MAX : Displayed  when  the

maximum  reading  is
shown 

MIN : Displayed  when  the

minimum  reading  is
shown 

ALARM:  :  Displayed  during

alarm  setting  and  blinks
during  alarm 

 
 
 
 
 
 
Displayed  when  temperature
readings are shown 
 
Displayed  when  emissivity
value is shown 
 

■Setting Emissivity (reference) 

The table below shows emissivity for some materials. Please note that these values 

are just for reference and only apply to IRT-100-TA. Infrared radiation depends on 

materials, condition of the surface and ambient temperature. 

 

Material 

Emissivity 

Material 

Emissivity 

Water, Ice 

0.98 

Cloth, Fabric 
(colored) 

0.95 

Soil 

0.92 to 0.96 

Leather, Fur 

0.96 

Concrete (wet) 

0.96 to 0.98 

Human skin 

0.99 

Concrete (dry) 

0.91 to 0.95 

Vegetable, Fruit 

0.98 

Ceramic   

0.85 to 0.95 

Dough 

0.98 

Plaster, asbestos 

0.92 

Meat 

0.98 

Plastic 

0.90 to 0.95 

Copper oxide 

0.5 to 0.6 

Rubber (black) 

0.95 

Ferro-oxide 

0.7 to 0.8 

Wood 

0.98 

Painted surface 

0.8 

Paper 

0.92 

Tile 

0.8 

 

Estimating emissivity using the Black Body Tape 

Place a piece of the black body tape (emissivity: 0.94) on an object. Set emissivity to 

0.94 in FREE mode and measure the temperature of the object. Then remove the tape 

from the object. Measure its temperature again and adjust emissivity value so that 

readings match the temperature measured with the black body tape attached on the 

object.  The  adjusted  value  is  the  emissivity  of  the  object.  The  black  body  tape  is 

available as an option. 

 

Measure Key 

Measurements are taken while
this  key  is  pressed.  Releasing
this  key  freezes  the  last
reading and automatically turns
the  instrument  off  in  30
seconds, if no key is pressed. 
 

Mode Key 

Display  the  current,  the
maximum  or  the  minimum
reading,  and  selects  setting
mode. 
 

Up Key 

Selects  DARK,  BRIGHT  or
FREE, and sets emissivity and
alarm value. 
 

Down Key 

Sets  emissivity  and  alarm
value. 
 

Entry Key 

Registers  set  values  and
switches 

the 

instrument

setting mode. 

Brightness Sensor 

Automatically  lights  up  the  display
back-light when it becomes dim. 